5. Operating troubles, causes and remedial action 
(Troubleshooting) 

 

No. 

Operating troubles 

Causes and remedial 

action 

 

Pump 
does 
 not  
start

 

Pump 
does 
not 
prime 

Delivery 
is not  
ob-
tained 
or 
reduced

Pressure 
head is 
not 
obtained

 

Irreg-
ular 
pump 
delivery

Pump 
oper-
ates 
noisily 

Pump 
is 
leaky

 

Motor
gets 
too 
warm

Display 
show 
Error 
code

 

 

 

 

 

 

 

 

 

The screws of the pump-chamber 

maybe not tightened enough. Fix it!  

 

 

 

 

 

 

 

 

Check the direction of flow showed by 
the arrow on the pump, in case of 
wrong way, turn the pump head. 

 

 

 

 

 

 

Check suction pipeline and TC- seals 
for tightness. 

 

 

 

 

 

Check suction head-increase suction 
line cross section. 

 

 

 

 

 

 

Check viscosity of liquid pumped. 

 

 

 

 

 

 

 

Check pump speed. Control speed of 
drive motor. Check voltage and 
frequency and the fuse  (6,3A T in the 
control box) 

 

 

 

 

 

 

Avoid air inclusions in the liquid to be 
pumped. 

 

 

 

 

 

 

 

Check pressure head-open valve in 
discharge line completely, remove 
obstruction in discharge line. 

 

 

 

 

 

 

 

 

Pressure line completely or partly 
clogged, diaphragm maybe broken, 
change diaphragm! 

10 

 

 

 

 

 

 

 

 

The diameter of the  pipes in suction or 
pressure line are too small. 

11 

 

 

 

 

 

 

 

 

Check the coupling halves. They must 
be fixed with 1mm space. 

12 

 

 

 

 

 

 

 

 

Check longitudinal play of coupling rod 
pins. The spider might be worn. 

13 

 

 

 

 

 

 

Check whether foreign bodies in pump. 
Disassemble pump, remove foreign 
bodies, replace defective parts. 

14  X 

 

 

 

 

 

 

Pump stopped by the thermal circuit 
breaker. Please allow the motor to cool 
down – please reduce the power 
consumption. 

15  X 

 

 

 

 

 

 

 

Bearings are worn or defective 

Disassemble pump, replace the shaft – 
bearing – cap unit (PSKITWLC155) 

16 

 

 

 

 

 

 

 

 

The valves are dry (e.g. not in use for a 
long time), deformed or worn. Change 
valve or wet the pump. 

17 

 

 

 

 

 

 

 

 

The diaphragm is burst ( the discharge 

pressure was too high) – replace it 
PSKITQ15. 

18 

 

 

 

 

 

O–rings between valve plate and pump 
housing are defective PSKITQ12.

 

19 

 

 

 

 

 

 

 

 

Align coupling accurately 

20 

 

 

 

 

 

 

 

The clamping ring screw got loose –

fixe it! 

21  X 

 

 

 

 

 

 

 

Parameters of the control panel are 

wrong – check basic settings. 

22 

 

 

 

 

 

 

 

 

Pump after SIP cooled down too fast – 

slow cooling with room temperature 

Содержание 150s

Страница 1: ...trol pa manua are of a ope At d the w Orig and tructi 50S Qu Piston ultiple U anel and hou l carefu an appr rating s ttentio warning inal Ins Insta ions aterna Diaph Use using other v ully befo ropria...

Страница 2: ...tructions 8 3 5 Safety hints for the operator 8 3 6 Safety instructions for maintenance inspection and mounting operations 9 3 7 Unauthorized reconstruction and spare part usage 9 3 8 Inadmissible mod...

Страница 3: ...nd mechanical influences The following storage conditions are recommended Steady ventilated dust and vibration free storage room Ambient temperature between 15 C 59 F and 25 C 77 F with a relative hum...

Страница 4: ...cosmetic research centers or plants Examples Solutions containing proteins albumin IgG Clotting factors monoclonal antibodies enzymes vaccines Solutions of polymers or suspensions silicon latex chroma...

Страница 5: ...er Prior to each use we recommend to flush the pump with a proper fluid e g water or buffer Prior to the very first use it might make sense to clean and sanitize the pump chamber A commercial caustic...

Страница 6: ...r Panasonic 50 W 230V with integrated frequency inverter Speed control Panasonic Digital Control Panel 2 5 Cleaning CIP Depending on the products in contact with the pump and given requirements the cl...

Страница 7: ...max 30 min Follow instructions of the autoclave manufacturer Please make sure that the pump chamber is not placed on the clamp ring during the autoclave process The clamp ring should not be loaded dur...

Страница 8: ...strictly the safety guidelines of this manual as well as all national and possible internal regulations e g the handling of chemicals like caustic or acid the handling of biological materials the han...

Страница 9: ...art up and check the tightness of the pump o Please respect the relevant additional security advices if the pump has been used for aggressive dangerous or toxic liquids e g suitable protective equipme...

Страница 10: ...he right tool for the specific application The user operator is responsible to perform a proper method of testing if the pump should be applied for his specific application The chemical and thermal co...

Страница 11: ...led personnel Check the electrical cables before connecting to mains supply During all maintenance work it has to be ensured that no explosive atmosphere can arise Appropriate protection equipment is...

Страница 12: ...ecessary to shorten the maintenance intervals for the elastomers significantly In case that the diaphragm broke it needs to be replaced Then it is also recommended to check the ball bearings For corro...

Страница 13: ...and the valves can be conducted by the operator The following drawings describe it step by step maintenance kit PSKITQF15MU In case of bursting diaphragms by overpressure we advice also to change the...

Страница 14: ...05 page 14 of 19 4 1 3 Assembly of the pump chamber to the ring drive Pay attention to the following torque values Image 2 item 1 Bolts pump chamber 4 Nm 3 lb ft Image 2 item 2 Bolt clamping ring 2 N...

Страница 15: ...the bearing unit shaft bearing cap unit kit PSKITWLC155 can be conducted by the operator The following drawings describe it step by step Between both coupling halves has to be a gap of 1 mm 0 039 inc...

Страница 16: ...truction in discharge line 9 X Pressure line completely or partly clogged diaphragm maybe broken change diaphragm 10 X The diameter of the pipes in suction or pressure line are too small 11 X Check th...

Страница 17: ...p Pump rpm motor rpm The flow rates showing at the chart above were ascertained with new diaphragms and new valves under standard conditions after the final assembly of the pump 0 0 5 1 1 5 2 2 5 3 0...

Страница 18: ...face approx cm inch 112 17 112 17 Speed range pump rpm 10 3000 10 3000 Connection specification inlet standard Connector 1 4 TC 1 4 TC Flange diameter mm inch 25 25 Internal diameter mm inch 4 7 4 7 C...

Страница 19: ...TDK Lambda power supply Designation n a HWS Analog input optional 0 5V 0 10 V Protection class IP54 IP20 switch board Power supply 230V 50Hz 1P optional 110V 60 Hz 1P 85 265VAC 50 60 Hz 8 2 A Housing...

Отзывы: