Foot-pedal and Hand-switch:
•
Dry Contact (closed contact) remote option, connect the black and red wires to anything that can close a contact and
short these 2 wires together.
•
1-30V AC/DC remote option, connect the white and the green wires to the remote signal. RESPECT THE INDICATED
POLARITY WHEN MAKING THIS CONNECTION.
DO NOT EXCEED 30V AC/DC.
To enable the remote system startup for either remote option, first plug the system into power and then connect remote
cable. Then simply have the other piece of equipment close (short) the dry contact or supply a remote control voltage.
When the unit is stopped the QUATRO System will shut down after a short delay.
After Remote Sensing Has Been Activated:
•
If the unit is RECEIVING a remote signal: System ON and Remote ON Lights will be FULLY Illuminated
•
If the unit is WAITING FOR a remote signal: System ON will be OFF and Remote ON Lights will be FLASHING
Unplugging the power cord wll automatically deactivate Remote Sensing Feature.