5. Maintenance and Troubleshooting
No daily maintenance is required. This equipment should only be serviced by
authorized technical personnel. Use only original spare parts and consumables. In
case of minor problems refer to the following:
Symptom
Inspection Procedure
Possible Solution
The air is flowing out from the
control
unit but the mattress is not inflating
•
Do the air tubes allow smooth
airflow? Is an air tube kinked?
•
Is there any air leakage from the
air cells?
•
Is there any air leakage from air
tube between mattress and
control unit?
•
Has the air tube connector been
connected properly?
•
Adjust the air tubes to enable
smooth air flow.
•
Replace with new air cells
•
Replace with new air tubes
•
Re-connect the air tubes.
The Control Unit is not working
•
Check the fuse
•
Replace with a new fuse.
•
Fuses may only be exchanged
by qualified and authorized
personnel.
Some air cells have abnormal low
air pressure while the air pressure
for other air cells is normal
•
Is the connection between air
cells and the manifold kinked?
•
Is there any air leakage from the
air cells?
•
Adjust the connection between
cells and manifold
•
Replace with a new air cell.
6. Warranty
Quart Healthcare Products guarantees that this equipment is free from defects in
materials and workmanship. Our obligation under this warranty is limited to the
repair of equipment returned to the place of purchase within 24 months of delivery
date
We agree to service/adjust any equipment returned, and to replace or repair any
part that is proven to be a warranty defect, at no charge
This warranty excludes equipment damage through shipping, tampering, improper
maintenance, carelessness, accident, negligence or misuse, or products that have
been altered, repaired or dismantled other than with the manufacture’s written
authorization and by its approved procedures and by properly qualified technicians
In no event shall Quart Healthcare Products be liable for any direct, indirect or
consequential damages or losses resulting from the use of equipment