Sleep Mode (If Enabled)
The power chair controller features a sleep mode. Sleep mode is a built-in circuit that will automatically shut off the main power
if the joystick is not moved in any direction for a period of time. This time factor is programmed into the controller. To restore
power and continue, push and key on the keypad.
Trouble Codes
The system status light is displayed within the on/off key. This light is lit if the system is turned on. It also flashes in groups
called flash codes, to indicate system faults. The table below identifies the individual fault codes. If your keypad displays one
of these codes, contact your authorized Pride provider.
FLASH CODE
SEQUENCE
DIAGNOSIS
SOLUTION
•—•—•—
DX module fault
See your authorized Pride provider.
••—••—••—
DX accessory fault
See your authorized Pride provider.
•••—•••—
Left motor (or connection) fault
Check left motor wiring.
••••—••••—
Right motor (or connection) fault
Check right motor wiring.
•••••—•••••—
Left park brake fault
Check motor/brake wiring.
••••••—••••••—
Right park brake fault
Check motor/brake wiring.
•••••••—•••••••—
Low battery fault
Check that the battery wiring is secure.
•••••••—•••••••—
Over voltage fault
Check that the battery wiring is secure.
•••••••—•••••••—
CANL fault
See your authorized Pride provider.
•••••••—•••••••—
CANA fault
See your authorized Pride provider.
•••••••—•••••••—
Stall timeout
Turn unit on, then off.
•••••••—•••••••—
Module mismatch
See your authorized Pride provider.
NOTE: You must turn the controller off and then on again to reset the controller, even if the source of the fault
is removed/corrected.
In the event of a fault, the system status light displays diagnostic indications. Faults are encoded as follows: one (for a low
battery voltage fault) to twelve (module mismatch) and are displayed by the light flashing the number of times prescribed by
the fault code. The flash sequence (one to twelve) is followed by a long off period (2 seconds). If more than one fault exists,
then the fault having the highest priority is indicated. The controller must be turned off and then on again even if the source of
the fault is removed. If you cannot resolve the problem, contact your authorized Pride provider.
OPERATION