MINISYSTEM Manual Version 1.5
Page 31
LS
#,#,###,###,###,###
LS?
Laser Setup. This is a combined command to adjust multiple
system parameters at once instead of issuing multiple
commands.
First Parameter = Select wavelength (1 = IR, 2 = GRN, 3 = UV, 4
= DUV),
Second Parameter = Selects High (1) or Low (0) mode,
Third Parameter = Set Energy Density (0
– 1000),
Fourth Parameter = Set polarization angle (-900 to +900),
Fifth Parameter = Set Slit X-axis (0.00 to 2500 microns, depends
on magnification),
Sixth Parameter = Set Slit Y-axis (0.00 to 2500 microns, depends
on magnification),
LS? returns current settings.
PA #
PA?
Polarization Angle. Sets the angle of the laser output polarization.
Units are 10*degrees. Values range from -900 to 900 for +/-90
degrees, where 0 is vertical. Default is 0.
PM #
PM?
Mode. Sets the laser firing mode, 0 = continuous, 1 = single shot,
2 = burst. (Default: 0). A query returns the currently set mode.
RR ##
RR?
Repetition Rate. This sets the rate at which the laser will fire at.
Values are from 1-50 Hz. A query will return the current rate.
SC?
Shot Count. Queries the laser accumulated shot count.
SW #
SW?
Select Wavelength. Selects the wavelength to control. Range is 1
to 4 where 1 = IR, 2 = GRN, 3 = UV, 4 = DUV. A query returns
the current wavelength being controlled. NOTE: The system will
only have two available wavelengths.
SY #
SY?
Sync Output Mode. 0 = Off, 1 = LD Pulse, 2 = Emission
UC 0
UC?
User Count. This is a user re-settable shot counter. It is a saved
count and can be cleared by issuing a 0 parameter following the
command.
Slit Commands
SA #
SA?
Slit Rotation. Sets the current slit rotation angle. Range is from
+45 to -45 degrees. A plus or minus is required to determine
angle. (Default: 0). Parameter = degrees*10 (+450 to -450). A
query returns the current rotation angle.
SP #,##.##,##.##
SP?
Slit Position. First Parameter is Magnification. Sets the current
slit X and Y blade position. Values are from 0.00 to 2500 microns
for both axes depending on magnification. Actual allowed range
depends on the objective magnification. Slit range is 0 to 2.5mm,
thus the image size range is 0 to 2500/magnification. A query
returns the current cut size.
Illuminator Commands