CubiScan 125 Command Set
Communications Protocol
94
CubiScan 125
Scale Calibration
This function is required when the load cell is replaced, when a new
controller is installed, or for routine calibration. This command
causes the CubiScan 125 to reply with a scale calibration code. This
command must be issued THREE times to complete scale calibra-
tion. Scale calibration codes are defined below.
Pos
Len
Description
Type
Range
ASCII
Command Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
6
Test Weight Value
Numeric
050.00-100.00
9
1
End of Text
Control
(ETX)
03h
10
1
Carriage Return
Control
(CR)
0Dh
11
1
Line Feed
Control
(LF)
0Ah
Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
1
Acknowledge
Alpha
(A)
41h
4
2
Identifier
Numeric
(00-99)
6
1
End of Text
Control
(ETX)
03h
7
1
Carriage Return
Control
(CR)
0Dh
8
1
Line Feed
Control
(LF)
0Ah
Negative Acknowledge Format
1
1
Start of Text
Control
(STX)
02h
2
1
Scale Calibration
Alpha
(S)
53h
3
1
Neg. Acknowledge
Alpha
(N)
4Eh
4
1
End of Text
Control
(ETX)
03h
5
1
Carriage Return
Control
(CR)
0Dh
6
1
Line Feed
Control
(LF)
0Ah
Содержание CubiScan 125
Страница 8: ...Table of Contents iv CubiScan 125 Notes...
Страница 20: ...Assembling the CubiScan 125 Setup 10 CubiScan 125 Figure 4 CubiScan 125 Base Assembly on Optional Cart...
Страница 39: ...Setup Connecting to a Computer or Network CubiScan 125 29 13 Click the Hardware tab...
Страница 44: ...Setup Checklist Setup 34 CubiScan 125 Notes...
Страница 90: ...Computer Error Messages Troubleshooting 80 CubiScan 125 Notes...
Страница 124: ...Parts List 114 CubiScan 125 Notes...