12 Working with optical fibers
Quantifi Photonics products are equipped with high quality optical connectors in compliance with EIA-455-21A standards.
CAUTION
Keep connectors clean and in good condition to ensure maximum power and to avoid erroneous readings:
Always inspect fiber end faces for cleanliness using a fiber inspection probe before inserting them into a port..
If required, clean fibers and faces as detailed below.
Quantifi Photonics is not responsible for damage or errors caused by bad fiber cleaning or handling.
NOTE
To avoid damaging ferrules or fiber faces due to mismatched connectors, always check ports and connector type information before inserting a
connector. All Quantifi Photonics units are labeled with connector type information.
When connecting a fiber-optic cable to a port:
1. Visually inspect the fiber end face using a fiber inspection microscope.
2. If a
connector end face
is dirty:
Wipe the connector end face using a reel-type cleaner and inspect again.
For stubborn hard to clean connectors:
l
Use lint-free fiber-cleaning wipes soaked in a fiber optic cleaning solution.
l
Wipe the connector on the soaked part.
l
Dry the connector by wiping on the dry part of the wipe, or by using a reel-type cleaner.
Repeat the process until connector inspection shows a clean fiber face.
3. If a
bulkhead inner connector face
is dirty:
Use a pen-type dry cleaner, align the cleaning tip with the port and push the cleaner until you hear the characteristic click. Inspect again.
For stubborn hard to clean bulkhead connectors:
l
Use a stick-type cleaner dipped in a fiber optic cleaning solution.
l
Carefully align and insert the stick into the connector and gently rotate the stick for several seconds applying light pressure.
l
Use a pen-type cleaner to dry the connector.
Repeat the process until connector inspection shows a clean fiber face.
4. If the fiber end face is clean:
Carefully align the connector and port to prevent the fiber end from touching the outside of the port or other surfaces.
If the connector features a key, mate it correctly into the corresponding notch of the port bulkhead.
Quantifi Photonics | SLED 1000 / 1300 Series PXIe User Manual | Document version 1.08
54 / 61