By example, we have checked the values on a device, at cold laser and at
warm laser.
For 1 pulse :
- at cold laser, we measured 540 V.
- at warm laser, we measured 565 V.
To get these values, the P1 potentiometer has been adjusted at nearly
618V.
For 2 pulses : - at cold laser, we measured 760V.
- at warm laser, we measured 765 V.
To get these values, the P2 potentiometer has been adjusted at nearly
780V.
For 3 pulses : - at cold laser, we measured 870V.
- at warm laser, we measured 910V.
To get these values, the P3 potentiometer has been adjusted at nearly
910V.
7-4 DISPLAY CONTRAST ADJUSTMENT
To improve the display contrast, you must use the R27 potentiometer on
the CPU Board.
R27 Potentiometer
7 - ELECTRONIC ADJUSTMENTS