SERVICE MANUAL
Rev. 05/2005
P - 38
5-1 ACCESSING SERVICE MODE
Accessing the SERVICE mode is only possible after opening the laser
console.
To access at the service mode :
- Insert jumper connector on the CPU board,
- Turn On the key : the laser is in service mode.
5-2 SCREEN IN SERVICE MODE
When the OPTIMIS II is in service mode, the energy value is measured by
the cell of the energymeter board and displayed in real time on the
OPTIMIS II screen.
You can check if the measured energy is in the safety windows (+20%, -
20% and -50%).
Screen in service mode, when you pulse :
Sub-D connector to
connect a PC computer
(linkage RS232)
Place here the
jumper connector
1.6mJ 1 E1.6mJ
Count 0
Energy selectionned by
the energy knob
Number of pulses
Counter value
Energy value read by
the energy meter board,
(displayed in real time)
5 - SERVICE MODE