background image

QA402 User Manual 

 

29 

 

 

If the measured response is within the mask boundaries, you will see PASS displayed, and FAIL display if not. 
Masks can be as complex as required to capture the expected behavior. You can create them manually, as 
above, or use an automated measurement to export a mask. 

Full Scale Input 

Full Scale Input determines the maximum input before clipping will occur. Note that the maximum input to the 
analyzer is +32 dBV, with a max DC of 40V and a max AC

PEAK

 + DC of +/- 56V. Those limits must always be 

respected, even if you are on the +42 dBV input range. The concept of full-scale input is to help the engineer 
understand how far away they are from clipping, as this has an impact on the measurement quality.  

The full-scale input control group is shown below: 

 

Note that for level 24 dBV and greater, the attenuator is active. The attenuator is a hardware divider at the 
front-end of the analyzer that attenuates the reduces the signal by 24 dB. When the input attenuator is active, 

Содержание QA40 Series

Страница 1: ...QA40x User Manual Revision 1 20 February 2022...

Страница 2: ...esponsible for shipping to us and we will be responsible for shipping it back to you Devices that have suffered a failure due to operation in excess of specified parameters can usually be repaired for...

Страница 3: ...A40x Analyzer Features 6 Analyzer Front Panel 6 LEDs 6 Expansion Connector 7 BNC Inputs and Outputs 7 Understanding Differential Measurements 8 Rear Panel Summary 8 Electrical Characteristics of the C...

Страница 4: ...n Settings 19 Weighting 20 Windowing 21 Measurements 21 Generators 23 Sine Generation 24 Multitone 25 White Noise 27 Expo Chirp 27 Full Scale Input 29 Run Stop 30 Run Stop Context Menu 31 Cursors 32 O...

Страница 5: ...shares a ground with the PC which in turn shares a ground with the power distribution in your office The audio side of the QA40x analyzer is isolated from the USB ground it is floating When you conne...

Страница 6: ...ht Outputs 24 bit ADC and DAC for QA402 and 32 bit ADC and DAC for QA403 Fully Isolated from the PC 8 Input Gain Stages from 0 dBV to 42 dBV full scale the maximum input is 32 dBV 4 Output Gain Stages...

Страница 7: ...ther than 48 kSPS are not currently supported but may be added in the future More information and schematics for use with the front panel expansion connector can be found on the QA40x BNC Inputs and O...

Страница 8: ...ases Rear Panel Summary The rear panel has a single USB connector This is designed for high speed 480Mbps USB connections The device consumes between 700 and 900 mA during normal operation The device...

Страница 9: ...Windows Version The QA40x application was developed on Windows 10 although it might run on previous versions of Windows USB Drivers On Windows you do not need to install drivers The hardware uses spe...

Страница 10: ...onfirmation as shown in the lower left corner After the device has been connected you can begin making a measurement by pressing the RUN button in the Run Stop control group Button and Control Panel O...

Страница 11: ...p that only permits a single selection the button that is currently illuminated will momentarily flash to signify it is turning off in response to your request to change modes QA40x Basic Controls The...

Страница 12: ...er Manual 12 Measurement Display Area At the top of the trace display is the measurement display area Measurements can be added by clicking in this region Clicking will bring up the Add Measurement di...

Страница 13: ...QA402 User Manual 13 You can filter on measurements by entering text in the Filter Measurements dialog For example if you enter thd case insensitive the test choices will collapse to the following...

Страница 14: ...y right clicking on a peak The markers are always sorted from highest amplitude and the tabular display will convey both absolute value and relative value referenced to marker 0 the highest peak Marke...

Страница 15: ...ent the latency compensation will need to be increased to ensure the signal is properly captured see Edit Settings Mirror Indicates Generator 1 settings are being mirrored to the PC s primary audio de...

Страница 16: ...nd hold After 1 second the menu will open This is useful if you are using a finger on a touch screen 2 Mouse right click The menu will immediately open A sample control stack is shown below The heavy...

Страница 17: ...specify The context menu for the dBV button is shown below In this dialog you can specify the units for the Y axis Normally dBV is used But you can also specify dBu You can also specify if you have a...

Страница 18: ...Or you could specify the level in dBr This is useful if for example you know that a mic should generated a certain output at a given sound pressure level When using the dBr setting you can also overri...

Страница 19: ...he averaging button will increase or decrease the average count by one If you click while pressing the CTRL button the count will change by 5 counts When averaging you might wish to re start the avera...

Страница 20: ...you apply standard A Weighting or a custom User weighting function The latter can be useful for flattening an RIAA response curve notching out undesired frequencies or compensating for known DUT defic...

Страница 21: ...ements You can quickly select the types of measurements you d like to make from this control group The buttons are one shot in that they don t stay lit If you press GAIN and you don t currently have a...

Страница 22: ...ecify your preference for THD and THD N measurements If you select dB then when you add a THD measurement it will default to dB Otherwise it will default to percent The fundamental selection is where...

Страница 23: ...group we can select one of four modes of operation Sine Gen Multitone White Noise and Expo Chirp Overall the QA40x has 5 modes of operation as shown in the block diagram below The fifth mode of opera...

Страница 24: ...exponentially swept in frequency from 20 Hz to Nyquist The exponential nature means you spend more time at lower frequencies and less time at higher frequencies But your energy per Hz is the same Sin...

Страница 25: ...sult in changes that step 1 kHz 2 kHz 3 kHz etc Multitone The multitone context menu is shown below Here you can specify the number of tones per octave For example if you specify 1 tone per octave the...

Страница 26: ...ines however is around 0 dBV But the real challenge is in the time domain Notice that even though each sine is just 100 mV in amplitude the total output level has exceeded 4V peak in places Be aware o...

Страница 27: ...g is done when windowing is non zero 1 A complex division is done between the output waveform over the input waveform This yields the frequency response 2 A reverse FFT is done on the complex division...

Страница 28: ...act on the frequency See more discussion on this topic at the QA40x Wiki If you have a mask file prepared you can load the mask file and opt to display a pass or fail message if the frequency response...

Страница 29: ...maximum input to the analyzer is 32 dBV with a max DC of 40V and a max ACPEAK DC of 56V Those limits must always be respected even if you are on the 42 dBV input range The concept of full scale input...

Страница 30: ...sample rate changes In general the performance of the converters is better at lower sample rates If your measurement task doesn t require frequencies beyond 20 kHz or so stick to the 48K sample rate...

Страница 31: ...r Manual 31 Run Stop Context Menu The context menu for the Run Stop button is shown below Triggered Acquisition The Triggered Acquisition is discussed in more detail on the QA40x Wiki article Triggere...

Страница 32: ...FS signal will be 18 dBV And if you specify the 8 dBV output range and a 20 dBFS volume then the max output will be 12 dBV Cursors The cursor control block is shown below You can add one or two cursor...

Страница 33: ...QA402 User Manual 33 The cursors also work in the time domain and the legend will also show 1 period allowing you to quickly learn the frequency at a given point in time...

Страница 34: ...FT you can emit a single short burst of sine For example a 1K FFT at 48Ksps will result in a 20 mS burst of signal Emitting a single one of these bursts can be helpful as you are trying to learn the l...

Страница 35: ...een from the webpage linked below while the QA40x application is running When you click on this link the QA40x application will generate a web page of all available tests including descriptions of eac...

Отзывы: