MH3418-01 

                                                                                                                                   

EN-4

 

 

  The Quanta Pulse Oximeter and its accessories should be disposed of properly. 

 

Battery Caution:

 There is a risk of explosion if the battery for the Quanta Pulse 

Oximeter is replaced by an incorrect type. 

  The Quanta Pulse Oximeter is intended for adult use. 

  Quanta  Pulse  Oximeter  does  not  provide  with  a  low  SpO2  alarm  condition  or 

pulse rate physiological alarm condition. 

  Quanta Pulse Oximeter does not support low perfusion detection. 

  Clean the skin of Quanta Pulse Oximeter sensor and charger with 75% alcohol; 

and clean the finger sleeve in wash machine. 

  If  you  wear  Quanta  Pulse  Oximeter  too  long,  it  may  cause  uncomfortable  by 

long  time  pressure.  Please  check  if  you  have  any  on  the  finger  every  eight 
hours. 

  Quanta  Pulse  Oximeter  Sensor  is  a  compact  all-in-one  design  with  Emitter  / 

Receiver embedded inside. So that it doesn’t have to verify the compliance of 
monitor, probe extenders and cable before use. 

  Quanta  Pulse  Oximeter  has  been  validated  and  tested  for  compliance  with 

international standard ISO 80601-2-61. 

  The self life of Quanta Pulse Oximeter is 2 years. 

  Quanta Pulse Oximeter is calibrated to display functional oxygen saturation. 

  The pulse rate accuracy is verified by the root-mean-square(rms) method which 

is the difference between pulse rate data recorded by Quanta Pulse Oximeter 
and with pulse rate data set by the SpO2 simulator.   

  Do  not  use  any  unauthorized  oximeter  simulator  to  assess  the  accuracy  of 

Quanta Pulse Oximeter. And if there has any question please contact Quanta 
Customer Support.     

  The  SpO2  accuracy  validation,  Quanta  Pulse  oximeter  clinical  trial  was 

performed  in  accordance  with  ISO  80601-2-61  and  FDA  guidance.  The  study 
population  is  healthy  volunteers  who  consent  to  induced  hypoxia  and  arterial 
blood  sampling  more  than  200  data  points.  Participated  subjects  is  including 
Asian,  Caucasian,  African,  Male,  and  Female.  Their  ages  are  from  21  to  50 
years old. The statistically study result indicated that the SpO2 accuracy (

±

Arms) 

of Quanta Pulse Oximeter was less than 

±

3 % in the SpO2 range 70% ~100% 

complied with the accuracy requirement. 

  When the finger sleeve is not viscous, or it is recommended to use finger sleeve 

for a month if it’s used every day.   

 

 

 

 
 
 

Содержание Q-spo-wu1

Страница 1: ...MH3418 01 Version 3A Q Qu ua an nt ta a P Pu ul ls se e O Ox xi im me et te er r Model Q spo wu1...

Страница 2: ...ice with Your Smartphone 11 Performing Measurement 12 Viewing Real time Readings 13 Switching Measurement Mode 14 Viewing History 15 Editing User Profile 16 Clear Data 17 Reset the Device 17 Exit APP...

Страница 3: ...ervice technician or user should not stare at the light Do not clip Quanta Pulse Oximeter on edema or tender tissue Do not use the Quanta Pulse Oximeter in a flammable environment Quanta Pulse Oximete...

Страница 4: ...national standard ISO 80601 2 61 The self life of Quanta Pulse Oximeter is 2 years Quanta Pulse Oximeter is calibrated to display functional oxygen saturation The pulse rate accuracy is verified by th...

Страница 5: ...dicated App called Q SPO2 APP The device should be worn on the index finger middle finger and ring finger along with a finger sleeve and it s for adult use only The Quanta Pulse oximeter is sold to pr...

Страница 6: ...ts In order to properly use the Quanta Pulse Oximeter the following items are required The Qunata Pulse Oximeter Sensor Finger Sleeve A Bluetooth enabled Android Smartphone with Android version 5 0 or...

Страница 7: ...the following 1 With the charging port is facing outwards insert the bottom side of the device halfway into the bottom opening of the finger sleeve 2 Insert the upper side of the device into the uppe...

Страница 8: ...tach the Velcro strip and flip it over to another side of the finger sleeve so that you can access the pogo charging port 3 Align and attach the pogo pin of the charger with the charging port of the d...

Страница 9: ...ructions to download and install Q SPO2 APP on your smartphone Note In order to install the Q SPO2 app your smartphone will need at least 8MB of storage capacity available Creating User Profile The fi...

Страница 10: ...easure your heart rate after at least 5 minutes of resting Obtain a more accurate heart rate measurement performed by a trained professional 3 Tap Done to save the data The main screen appears Note Yo...

Страница 11: ...oth devices are within Bluetooth range 10m To pair the device with your smartphone do the following 1 Press the Power button for one second flash green LED to turn the device on 2 On your smartphone l...

Страница 12: ...make sure you have successfully paired your smartphone with the device 1 Insert your index finger fully into the finger sleeve Make sure your finger is accurately placed on the probe of the device No...

Страница 13: ...level and pulse rate would update every three seconds And it has no alarm functions The device will trigger smartphone to vibrate when the battery level is low If the current time displays and these...

Страница 14: ...of SpO2 and pulse rate By default the measurement mode is set to Continuous monitoring To switch to spot check measurement do the following 1 Tap 2 Tap Spot Check In spot check mode the sensor perfor...

Страница 15: ...e and scalable PPG waveform The information appears in History main screen as following Quick Search Slide the timeline scroll bar to quickly select the desired time The numbers on the scroll bar repr...

Страница 16: ...Tap or to go to other months Select the desired date and tap Done to view the recorded data on that day Selected date Today s date Recorded data Note Q SPO2 app will store the recorded data up to 7 da...

Страница 17: ...e to restore the factory default settings A confirmation message appears on the screen Tap OK to reset the device Exit APP Tap Quit to stop Q SPO2 app from running in the background A confirmation mes...

Страница 18: ...ing rate 52Hz G sensor accuracy 0 0156g Connectivity Bluetooth 4 0 BLE Transmit distance 10 meters open space USB USB2 0 Battery 3 7V 80mAh Battery Life Continuous measuring 5 hours Spot checking 7 ho...

Страница 19: ...MH3418 01 EN 19 Charger Specifications Item Specifications Input 5V 0 5A Operating Temperature 5 C 40 C Storage Temperature 20 C 60 C Weight 18g Dimensions 30 7 x 17 x 20 64 mm...

Страница 20: ...s within 10 meters SpO2 or pulse rate does not display on the main screen Finger might not be placed accurately on the probe Retry by inserting the finger into the finger sleeve until it touches the p...

Страница 21: ...a Customer Support Department Quanta Computer Inc QCI Address No 188 Wenhua 2nd Rd Guishan Dist Taoyuan City 333 Taiwan TEL 886 3 327 2345 FAX 886 3 318 4207 Email MedicalSales quantatw com EU Represe...

Страница 22: ...r charger is shown below In the serial number on the product label the 6th and 7th character starting from the left represents the year of manufacture while the 8th and 9th character represents the we...

Страница 23: ...that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Compliance Guidance and manufacturer s declaration ele...

Страница 24: ...po wu1 power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE UT is the a c mains voltage prior to application of...

Страница 25: ...gnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Страница 26: ...e communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GH...

Страница 27: ...der number Indicates the manufacture s serial number Indicates the manufacturer s name and address To indicate on the rating plate that the equipment is suitable for alternating current only to identi...

Страница 28: ...and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and r...

Страница 29: ...cteristic as well as performance to an approved low power radio frequency devices Article 14 The low power radio frequency devices shall not influence aircraft security and interfere with legal commun...

Отзывы: