![QMI 1000 Скачать руководство пользователя страница 30](http://html2.mh-extra.com/html/qmi/1000/1000_manual_3219630030.webp)
QMI 1000
manual
Page
30
10. ANALYSER SPECIFICATION
(NOTE MAY BE
SUBJECT TO CHANGE)
Parameter
Range
Resolution
Accuracy
Temp Measurement
Flue Temperature
Inlet Temperature
(Internal sensor)
Inlet Temperature
(External sensor)
0-600
o
C
0-50
o
C
0-600
o
C
0.1
o
C
0.1
o
C
0.1
o
C
+2.0
o
C
+0.3% reading
+1.0
o
C
+0.3% reading
+2.0
o
C
+0.3% reading
Gas Measurement
Oxygen
*2
Carbon monoxide
Carbon dioxide
Efficiency
*2
Excess Air
*2
CO/CO
2
ratio
*2
0-21%
0-2,000ppm nom
4,000ppm max
for 15 mins
0-30%
0-99.9%
0-250%
0-0.999
0.1%
1ppm
0.1%
0.1%
0.1%
0.0001
+0.2%
*1
+10ppm <100ppm
*1
+5% reading
+0.3% reading
+1.0% reading
+0.2% reading
+5% reading
Pressure (differential)
Nominal range +80mBar
Maximum over range
without damage to sensor is
+400mBar
+0.2 mBar
+1 mBar
+80 mBar
0.001 mBar
+0.005 mBar
+0.03 mBar
+3% of reading
Pre-programmed Fuels
Natural gas, Propane, Butane, LPG Light Oils (28/35 sec)
Storage Capacity
99 Combustion tests
20 Pressure tests
20 Tightness tests
20 Temperature tests
20 Room CO tests
*1
Using dry gases at STP
*2
Calculated
Содержание 1000
Страница 1: ...QMI 1000 Flue Gas Analyser with direct CO2 measurement Stock Number 18850 Dec 2009...
Страница 3: ...QMI 1000 manual Page 3 12 ELECTROMAGNETIC COMPATIBILITY 32 APPENDIX 1 MAIN PARAMETERS 33 35...
Страница 6: ...QMI 1000 Flue Gas Analyser with direct CO2 measurement...
Страница 36: ...QMI 1000 manual Page 36 Thank you for buying this instrument...