4.8
Calculating Vessel Centering
4.8.1
The actual vessel centering or spindle to vessel misalignment is
calculated by dividing the total indicated reading or T.I.R. by
two.
4.8.2
Example: The dial indicator moved clockwise from 40 to 15. The
total indicated reading is 1.5mm. Thus, 1.5mm divided by
two equals .75mm.
4.8.3
Example: The dial indicator moved clockwise from 40 to 25. The
total indicated reading is 2.5mm. Thus, 2.5mm divided by
two equals 1.25mm.
5.
Qualification and Validation
Prior to shipment, the kit is calibrated and includes Certificates of
Calibration and Traceability. The complete kit should be returned to QLA
for recalibration once a year.
6.
Maintenance
Although the measuring tools and gages contained in this kit are
maintenance free and do not require any daily service, it is
recommended that after each use, the gages and tools are returned to
the plastic protective case for storage.