Maintenance
PyroMark Q96 ID User Manual 01/2016
7-5
7.6
Cleaning and testing the reagent
cartridge
If the reagent cartridge is to be reused, it must be cleaned
directly after use. It is recommended that the reagent
cartridge is used a maximum of 20 times.
WARNING
Sharp needles
[W7]
Do not touch the sharp needles at the bottom of the
reagent cartridge. Handle the needles with care. Small
particles and fibers may obstruct the needles.
Note
: Be sure to observe all federal, state, and local
environmental regulations for the disposal of laboratory
waste.
The following items are required:
Powder-free gloves
High-purity water (Milli-Q
18.2 MΩ x cm or equivalent)
Clean, lint-free cloths
There are two possible cleaning procedures: A standard
cleaning procedure (Section 7.6.1) and a cleaning procedure
for clogged cartridges (Section 7.6.2). Ensure that a cleaned
cartridge is usable by performing a cartridge function check
(Section 7.6.3).
7.6.1
Standard cleaning procedure
Procedure:
1.
Discard any remaining solutions.
2.
Rinse the cartridge four times with high-purity water.
3.
Spray the outside of the needles with high-purity water.
4.
Fill the compartments to the top with high-purity water.
Hold the cartridge over a sink or beaker while pressing
firmly on top of each compartment with a finger (wear
powder-free gloves). A jet of water should come straight
out of the tip of the needle.
5.
If a needle is blocked, wet the needle tip with a drop of
high-purity water.
Содержание PyroMark Q96 ID
Страница 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Страница 2: ......
Страница 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 32: ...Installation Procedures 4 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 58: ...Operating Procedures 5 26 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 125: ......
Страница 126: ...1100446 01 2016 Sample Assay Technologies...