Operating Procedures
5-6
PyroMark Q96 ID User Manual 01/2016
Instrument parameters
In the “Instrument Parameters” dialog box, the following
parameters are available.
Reagent pressure
Pressure (millibar) for
dispensation of the enzyme
mix and substrate mix.
Enzyme pulse time
Dispensation time
(milliseconds) for the enzyme
mix.
Substrate pulse time
Dispensation time
(milliseconds) for the
substrate mix.
Nucleotide pressure
Pressure (millibar) from the
dispensation of nucleotides.
Nucleotide pulse
time
Dispensation time
(milliseconds) for nucleotides.
Note
Note about the instrument
method (optional).
5.4
Sample preparation
Equipment and reagents to be supplied by the user
PyroMark Q96 Vacuum Workstation
PyroMark Q96 Plate Low (cat. no. 979002)
PyroMark 96 Sample Prep Thermoplate Low (supplied
with the PyroMark Q96 ID)
Mixer for immobilization to beads
Heating block capable of attaining 80°C
96-well PCR plate and adhesive foil or strip caps
Biotinylated PCR template
Streptavidin Sepharose
®
HP (GE Healthcare, cat. no.
17-5113-01)
Sequencing primer
High-purity water (Milli-Q
18.2 MΩ x cm or equivalent)
Ethanol (70%)
Содержание PyroMark Q96 ID
Страница 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Страница 2: ......
Страница 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 32: ...Installation Procedures 4 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 58: ...Operating Procedures 5 26 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 125: ......
Страница 126: ...1100446 01 2016 Sample Assay Technologies...