![Qiagen Hybrid Capture System Multi-Specimen Tube Vortexer... Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/qiagen/hybrid-capture-system-multi-specimen-tube-vortexer/hybrid-capture-system-multi-specimen-tube-vortexer_user-manual_3176677018.webp)
18
Hybrid Capture System MST Vortexer 2 User Manual 09/2015
5
General Operation
5.1
Continuous vortexing
Note
: The
digene
HC2 DNA tests only use the continuous vortexing operational mode.
1.
Move the red lever to the horizontal position.
2.
Position the specimen rack and lid on the vortexer platform until it fits snugly within the
guides. If using a Conversion Rack, place the specimen rack with the notched corner of
the specimen rack in the right-front position of the vortexer platform.
3.
To lock the specimen rack in place, push the red handle all the way down to the vertical
position.
4.
Turn the motor speed dial clockwise to the applicable setting.
5.
Make sure the pulser toggle switch is in the
OFF
position.
6.
To begin operation of the MST Vortexer 2, turn the power toggle switch to the
ON
position.
The MST Vortexer 2 speed will be consistent until the power toggle switch is set to
OFF
.
5.2
Pulsating vortexing
1.
Move the red lever to the horizontal position.
2.
Position the specimen rack and lid on the vortexer platform until it fits snugly within the
guides. If using a Conversion Rack, place the specimen rack with the notched corner of
the specimen rack in the right-front position of the vortexer platform.
3.
To lock the specimen rack in place, push the red handle all the way down to the vertical
position.
4.
Turn the motor speed dial clockwise to the applicable setting.