14
Hybrid Capture System MST Vortexer 2 User Manual 09/2015
3
Installation
3.1
Unpacking
Before using the MST Vortexer 2 for the first time, examine the exterior carton and the
equipment itself for damage. In the event of shipping damage, contact your local QIAGEN
representative or QIAGEN Technical Services.
Carefully unpack the instrument and verify the contents of the package, which should
contain the following components of the instrument:
1 MST Vortexer 2 (base with platform)
1 power cord
If any of these items are missing, contact your local QIAGEN representative or QIAGEN
Technical Services. Save the original packaging until the instrument is operated successfully.
3.2
Start up
Plug the power cord into a grounded wall socket. The motor speed control electronics of the
MST Vortexer 2 depend on a sinusoidal power source. Do not connect the MST Vortexer 2 to
a UPS or other device that produces a stepped or square power waveform.