Troubleshooting
8-2
DML 3000 User Manual
8.3
“Door is Open” message
This error occurs when the access door is open. Close the access door and restart the
measurement of the microplate.
8.4
“Erratic Results” message
Possible causes
:
The DML 3000 is powered ON but has not warmed up
High background condition
If the DML 3000 is not powered ON at all times, allow the DML 3000 to warm up
for at least an hour prior to microplate measurement.
Measure the empty microplate mask.
8.5
Rattling noise or “Mechanical Failure” message
This error may be caused by the plate mask not being secure while the microplate is
in the DML 3000.
Power OFF the DML 3000.
Open the access door.
Reseat the microplate in the plate carrier.
Close the plate mask.
Power ON the DML 3000.
Click the “Cancel” button.
Remeasure the microplate.
8.6
Grinding noise or “Mechanical Failure” message
This error may be caused by the buildup of reagent residue on the plate mask, optical
mask, or guide rails. Clean the DML 3000. See “Cleaning the interior of the
DML 3000,” page 7-1, for additional instructions.
8.7
“Background too high, measurement cancelled” message
This error may be caused by the presence of fluorescing contamination, causing the
background to be out of range. Clean the DML 3000. See “Cleaning the interior of the
DML 3000,” page 7-1, for additional instructions.
Содержание DML 3000
Страница 5: ...Contents DML 3000 User Manual Contents 3 FCC Declaration D 1 Index Index 1 ...
Страница 6: ...Contents Contents 4 DML 3000 User Manual This page intentionally left blank ...
Страница 16: ...Functional Description 3 2 DML 3000 User Manual The following is the front view of the DML 3000 ...
Страница 34: ...Functional Description 3 20 DML 3000 User Manual This page intentionally left blank ...
Страница 42: ...Installation 5 4 DML 3000 User Manual Example of the DML 3000 with the HC2 System PC ...
Страница 43: ...Installation DML 3000 User Manual 5 5 Diagram of connecting a DML 3000 with a RCS ...
Страница 44: ...Installation 5 6 DML 3000 User Manual Diagram of connecting a DML 3000 with a LIS ...
Страница 45: ...Installation DML 3000 User Manual 5 7 Diagram of connecting a DML 3000 with a RCS and a LIS ...
Страница 46: ...Installation 5 8 DML 3000 User Manual This page intentionally left blank ...
Страница 54: ...General Operation 6 8 DML 3000 User Manual This page intentionally left blank ...
Страница 57: ...Maintenance DML 3000 User Manual 7 3 Example of wiping the interior of the plate mask ...
Страница 64: ...Maintenance 7 10 DML 3000 User Manual This page intentionally left blank ...
Страница 72: ...Appendix C C 2 DML 3000 User Manual This page intentionally left blank ...
Страница 74: ...Appendix D D 2 DML 3000 User Manual This page intentionally left blank ...
Страница 76: ...Index Index 2 DML 3000 User Manual Waste electrical and electronic equipment C 1 ...
Страница 77: ...Appendix D DML 3000 User Manual 3 This page intentionally left blank ...
Страница 78: ...Index Index 4 DML 3000 User Manual This page intentionally left blank ...
Страница 79: ...Appendix D DML 3000 User Manual 5 This page intentionally left blank ...
Страница 80: ...1058536CS Rev 01 04 2012 Sample Assay Technologies ...