![Qiagen DML 3000 Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/qiagen/dml-3000/dml-3000_user-manual_765619050.webp)
General Operation
6-4
DML 3000 User Manual
Visually make sure that the holes of the plate mask align with the microplate
wells.
WARNING
Risk of inaccurate test results
The access door must be completely closed during microplate measurement to
prevent ambient light from entering the DML 3000.
Ambient light could impact test results.
Close the access door.
6.3
Securing the DML 3000 using the transport locks
Use the transport locks to secure the internal components of the DML 3000 during
transport.
Loosen the red thumb screws of both transport locks.
Slide the transport lock on the optical head guide rail all the way to the right, and
tighten the thumb screw.
Slide the transport lock on the left guide rail to the back of the DML 3000, and
tighten the thumb screw.
View of the transport locks in the secured position:
Содержание DML 3000
Страница 5: ...Contents DML 3000 User Manual Contents 3 FCC Declaration D 1 Index Index 1 ...
Страница 6: ...Contents Contents 4 DML 3000 User Manual This page intentionally left blank ...
Страница 16: ...Functional Description 3 2 DML 3000 User Manual The following is the front view of the DML 3000 ...
Страница 34: ...Functional Description 3 20 DML 3000 User Manual This page intentionally left blank ...
Страница 42: ...Installation 5 4 DML 3000 User Manual Example of the DML 3000 with the HC2 System PC ...
Страница 43: ...Installation DML 3000 User Manual 5 5 Diagram of connecting a DML 3000 with a RCS ...
Страница 44: ...Installation 5 6 DML 3000 User Manual Diagram of connecting a DML 3000 with a LIS ...
Страница 45: ...Installation DML 3000 User Manual 5 7 Diagram of connecting a DML 3000 with a RCS and a LIS ...
Страница 46: ...Installation 5 8 DML 3000 User Manual This page intentionally left blank ...
Страница 54: ...General Operation 6 8 DML 3000 User Manual This page intentionally left blank ...
Страница 57: ...Maintenance DML 3000 User Manual 7 3 Example of wiping the interior of the plate mask ...
Страница 64: ...Maintenance 7 10 DML 3000 User Manual This page intentionally left blank ...
Страница 72: ...Appendix C C 2 DML 3000 User Manual This page intentionally left blank ...
Страница 74: ...Appendix D D 2 DML 3000 User Manual This page intentionally left blank ...
Страница 76: ...Index Index 2 DML 3000 User Manual Waste electrical and electronic equipment C 1 ...
Страница 77: ...Appendix D DML 3000 User Manual 3 This page intentionally left blank ...
Страница 78: ...Index Index 4 DML 3000 User Manual This page intentionally left blank ...
Страница 79: ...Appendix D DML 3000 User Manual 5 This page intentionally left blank ...
Страница 80: ...1058536CS Rev 01 04 2012 Sample Assay Technologies ...