FEET ADJUSTMENT
Screw the adjustable feet to the underside of the oven.
Do not use the oven without its feet.
Do NOT remove the feet.
The feet allow airflow to cool the oven,
preventing the oven from over-heating and damaging the
components.
The oven must be placed on a level, flat surface. Use a leveling tool to
verify a horizontal, level position. If needed, the feet can be adjusted
individually by screwing or unscrewing them as needed to achieve a
horizontal, level balance.
V. ELECTRICAL CONNECTION
WARNING: Connection to the power outlet must be conducted by a qualified and authorized
technician according to national, international and local laws in the country where installed.
Before connecting to the power outlet, ensure that the frequency (Hz) and voltage (V) meet the
requirements indicated on the information plate on the back of the oven.
WARNING
: if the oven does not come with a trolley to facilitate transport, it is recommended to
connect the oven first to the power outlet and then place the oven into position in the work area to
ensure the power cable is not damaged during placement. This process enables easy and safe
connection.
OVENS WITH CABLE AND SINGLE-PHASE PLUGS
Ovens with the
3xAWG14 15A Hospital Grade
cable and plug: the plug should be inserted into the
appropriate power socket. The product must be positioned to ensure the plug is always accessible.
Only the cable can be substituted. If substituting the cable, be advised the ground wire is yellow green.
Substitution of the cable must be completed by a qualified, authorized technician.