QIRF Operation And Maintenance Manual
QIRF Operation Manual
May. 01, 2012
85050-301-000 Rev C
22
3.10 Menu “3_Span Calibration”
3_ SPAN
CAL
SPAN CAL?
NN PPM
S3 R134a
CAL...
ACCEPTED
YES
ESC
YES
Span Calibration Flow Chart
3.10.1 Span Calibration Procedure
1. Connect tubing to QIRF
•
For Diffusion Version: Connect the cal gas supply tubing to the fitting on the
right side of enclosure
•
For Pump-through Version: Connect the cal gas supply tubing to the push-in
fitting on the bottom side of enclosure
2. Turn on the gas flow and press Key [YES] to display current reading
SPAN CAL?
NN PPM
This screen is displaying the current reading and asking if you want to perform
Span Calibration.
3. Waiting for about 3 minutes or till the reading is stable.
4. Press Key [YES] to perform Span Calibration.
5. During Span Calibration, the LCD will display the digital pot positions and
calibration statuses. It will take about 1 minute to perform Span Calibration, then the
span calibration data is saved and LCD displays “Accepted”.
6. If the LCD displays “Cal Error” that means something is wrong in the procedure,
repeat procedure 4 to try again. If the “Cal Error” is still displayed in the end, the unit
needs to be repaired in factory, otherwise, Span Calibration has succeeded, go to next
step.
7. Turn off the gas flow and remove it.