![Qardio QardioCore Скачать руководство пользователя страница 46](http://html2.mh-extra.com/html/qardio/qardiocore/qardiocore_user-manual_3100584046.webp)
46
Guidance and manufacturer’s declaration-electromagnetic emissions
The QardioCore Wireless ECG Monitor system is intended for use in the electromagnetic environment specified below.
The customer or the user of the QardioCore Wireless ECG Monitor system should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
CE emissions CISPR11
Group 1
The QardioCore Wireless ECG Monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RE emissions CISPR11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
The QardioCore Wireless ECG Monitor is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Not applicable
Содержание QardioCore
Страница 1: ...WIRELESS ECG MONITOR USER MANUAL...