The Pyxis ST-735 inline turbidity sensor can be inserted to the compression fitting port of Pyxis custom-
made ST-001 Tee

 

assembly.  The companion tee has two ¾ inch female NPT ports and can be installed to 

an existing ¾ inch sample water line.  The ST-735 can be also submersed into a vessel with using the 
Pyxis MA-102S submersion adapter. 

 

The ST-735 measures light scattering for the 0 

 100 NTU range using a white LED as the incident light 

source, the 100 

 1000 NTU range using a 950 nm LED light source, and the 950 nm absorbance for the 

1000 

 10,000 NTU range.  The ST-735 turbidity sensor automatically selects the measurement method 

needed for the sample according to the sample turbidity range detected when inserted inline. 

 

The ST-735 probe can be easily removed from the system without the need of any tools for being 
cleaned.  The Pyxis ST-735 probe has a short fluidic channel and is be easily cleaned.  The fluidic and 
optical arrangement of the ST-735 probe are designed to overcome shortcomings associated with other 
turbidity meters. 

 

The 4-20mA current output of the ST-735 probe is internally isolated.  It can be connected to any 
controller that accepts an isolated or non-isolated 4-20mA input.  

 

The Pyxis ST-735 probe uses long life LEDs as light source and will not require replacement in expected 
life unlike conventional turbidity probes.  This greatly enhances the robustness of the ST-735 probe.  It 
can be operated under a wide range of ambient conditions without the need for humidity and 
temperature regulation.  The performance of the ST-735 probe can be stable and consistent for a very 
long period time.   

2.

 

      Unpackaging the Instrument

  

Remove the instrument and accessories from the shipping container and inspect each item for any 
damage that may have occurred during shipping.  Verify that all accessory items are included.  If any 
item is missing or damaged, please contact Pyxis Lab Customer Service at 

[email protected]

 

2.1.

 

      

Standard Accessories Included 

 

ST-001 Tee Set

 

(tee, O-ring, and nut) 

 

Standard cable set with water proof connectors 

 

Instrument Manual, also available from 

www.pyxis-lab.com 

 

Содержание ST-735

Страница 1: ...1 ST 735 Inline Turbidity Sludge Concentration Sensor Operation Manual Version 1 1 December 2018...

Страница 2: ...n Screen and Reading 8 5 5 Device Screen 9 6 Probe Calibration with uPyxis Desktop App 9 6 1 Download uPyxis Desktop App 9 6 2 UnZip uPyxis Desktop App 10 6 3 Installing uPyxis Desktop App 11 6 4 Conn...

Страница 3: ...e ST 735 probe are designed to overcome shortcomings associated with other turbidity meters The 4 20mA current output of the ST 735 probe is internally isolated It can be connected to any controller t...

Страница 4: ...4 49 C Temperature Storage 20 140 F 7 60 C Sample Pressure 100 PSI Cable Length 5 feet terminated with IP67 connectors Power consumption 65 mA at 24 V Dimension Length 6 8 inch 172 7 mm body diameter...

Страница 5: ...ry to connect the 4 20 mA negative wire blue to the 4 20 mA negative terminal in the controller If a separate DC power supplier other than that from the controller is used make sure that the output fr...

Страница 6: ...ded to connect the ST 735 probe to your smart phone via Wi Fi Bluetooth interface 4 3 Connect via USB Figure 2 shows the connection between a computer and the ST 735 probe via USB RS485 adapter Use th...

Страница 7: ...pp Download uPyxis Mobile App from Apple App Store or Google Play 5 2 Connecting to uPyxis Mobile App Turn on Bluetooth on your mobile phone Do not pair the phone Bluetooth to the ST 735 Open uPyxis M...

Страница 8: ...ion Screen and Reading When connected Mobile App will default to the Calibration screen From the Calibration screen you can perform calibrations by pressing on Zero Calibration Low Range Calibration M...

Страница 9: ...Info Screen From the Device Info screen You can name the Device or Product 6 Probe Calibration with uPyxis Desktop App 6 1 Download uPyxis Desktop App Download uPyxis Desktop App from https pyxis lab...

Страница 10: ...10 6 2 UnZip uPyxis Desktop App Find your downloaded uPyxis Setup 1 3 8 file Right Click on the file Extract All and then Extract...

Страница 11: ...e the uPyxis Desktop App has been extracted Find the extracted uPyxis Setup file and left click click on Run and then click Install After install has been clicked the Setup Progress will continue Foll...

Страница 12: ...12 6 4 Connecting to uPyxis Desktop App Open uPyxis Desktop App on your desktop When the desktop app opens to find your device click on Device then Connect via Wi Fi...

Страница 13: ...List then the probe is connected Once connected to the probe the default main screen is the Information screen On the main screen you can set the information description for Device Name and Product Na...

Страница 14: ...e device click on Calibration On the Calibration screen there are three calibration tabs Zero Calibration Slope Calibration and 4 20 mA Span The screen does also display the reading of the device The...

Страница 15: ...15 6 8 Zero Calibration To perform Zero Calibration click on Zero Calibration Then follow the instruction on how to calibrate then click Ok...

Страница 16: ...16 6 9 Slope Calibration Low Range To perform Slope Calibration Low Range click on Slope Calibration Low Range Then follow the instruction on how to calibrate then click Slope Calibration...

Страница 17: ...17 6 10 Slope Calibration Mid Range To perform Slope Calibration Mid Range click on Slope Calibration Mid Range Then follow the instruction on how to calibrate then click Slope Calibration...

Страница 18: ...18 6 11 Slope Calibration High Range To perform Slope Calibration High Range click on Slope Calibration High Range Then follow the instruction on how to calibrate then click Slope Calibration...

Страница 19: ...ck Slope Calibration 7 Communicating using Modbus RTU The ST 735 probe is configured as a Modbus slave device In addition to the NTU value many operational parameters including warning and error messa...

Страница 20: ...ge at temperature over 100 F In an outbox installation properly shield the probe from direct sunlight and precipitation 8 1 ST 735 Inline Probe Cleaning Solution Soak the lower half of the ST 735 prob...