1. Power on the SP-710B by pressing
.
2. Press
System
(
) to launch the
DEVICE INFORMATION
screen.
3. Press
Diagnosis
(
) to launch the
SYSTEM DIAGNOSIS
screen.
4. Allow 5–10 seconds for the message in the top-right corner of the display change from
Starting BTLE...
to
BTLE Started
(Figure 32).
5. Choose to connect via one of two options:
(a) The
uPyxis®
Mobile App (see the
section), or
(b) The
uPyxis®
Desktop App (see the
section).
8.2.1
Calibrate an ST-500 Series Probe with the SP-710B via Bluetooth
The SP-710B can be used to verify the result of an inline Pyxis ST-500 Series probe by measuring the sample
water taken from the inline probe sample line. The SP-710B can then be used to calibrate the inline probes
over the Bluetooth connection. To calibrate an inline probe, follow the steps below:
1. Power on the SP-710B by pressing
.
2. Press
System
(
) to launch the
DEVICE INFORMATION
screen.
3. Press
Comm
(
) to launch the
COMMUNICATION
screen (Figure 36).
4. Press
Scan
(
) to begin scanning for Bluetooth devices.
5. Discoverable devices will begin to populate on the display with their name and MAC-Address (Figure
6. If more than one device appears in the
Device list
, press
»
(
) to cycle through the devices.
7. If no devices or the incorrect device appear in the
Device list
, press
Scan
(
) to re-scan for discov-
erable devices.
8. Press
Connect
(
) to begin pairing to the selected probe.
9. When the connection is established, the SP-710B displays the latest PTSA measurement from the
connected probe (Figure 38).
10. Fill the main module sample cell with the same sample water that the probe is measuring.
11. Press
Read
(
) to see PTSA measurement from the SP-710Balong with the probe measurement
(Figure 39).
12. Press
Calib
(
) to begin probe PTSA calibration.
13. The SP-710B will take the probe PTSA measurement three times to verify the calibration (Figure 40).
*
NOTE
*
It takes about one minute for the probe to approach the calibrated reading
and the three verifying readings may not be exactly the same as the value measured
by the SP-710B. Press
Read
(
) again to take more readings from the probe, if
necessary.
SP-710B User Manual
| +1 (866) 203-8397
25