2.

 

Starting SP-395 
 

2.1. 

Battery Installation 

 

The SP-395 is powered by a 9-volt alkaline battery. Do not use rechargeable nickel cadmium (NiCad) or  
lithium batteries. A typical 9V battery lasts for two months and enables about 3000 measurements.  
When the battery capacity is critically low, the SP-395 will display a LOW BATTERY warning for 5  
seconds and then automatically turn off.  
 
Replace the battery to resume operation of the SP-395 after the battery warning. The SP-395 will  
automatically turn on in the measurement mode after new battery installation.  
 
The SP-395 battery compartment, shown in Figure 2, is on the back side of the instrument. 
 
Install battery as follows:  
 

1.

 

Remove the battery compartment cover by loosening two screws. 

 

 

2.

 

Make sure that the smaller circular terminal (positive) of the battery is aligned with the  
hexagonal socket (positive) of the battery holder and the hexagonal socket (negative) of the  
battery with the circular terminal of the holder. Snap the battery firmly into the battery holder. 
 

3.

 

Replace the battery compartment cover, making sure that the sealing O-ring is lying flat on the  
battery holder. To prevent the SP-395 from accidently being turned on due to vibration, please  
firmly tighten the two screws.  

 

 

 
 

 

 

Содержание SP-395 HST

Страница 1: ...1 SP 395 HST Fluorometer Operational Manual Version 1 3 November 2018...

Страница 2: ...1 5 Optional Accessories 5 1 6 Light Shield Cover 5 2 Starting SP 395 6 2 1 Battery Installation 6 2 2 Description of Control Keys 7 2 3 Turning On Off SP 395 8 3 HST Measurement 8 3 1 Measurement 8...

Страница 3: ...Pyxis warrants that any labor services provided shall conform to the reasonable standards of technical competency and performance effective at the time of delivery All service interventions are to be...

Страница 4: ...ion is not required HST measurement is independent of sample pH in the range of 6 to 10 Extra color and turbidity parameters measured are used for automatic compensation to eliminate interference No s...

Страница 5: ...tion Guide 9V Alkaline Battery 3mL Transfer Pipet Full Instrument Manual in available from www pyxis lab com 1 5 Optional Accessories Carry Case for SP 395 1 ppm HST Standard 1 6 Light Shield Cover Fi...

Страница 6: ...t mode after new battery installation The SP 395 battery compartment shown in Figure 2 is on the back side of the instrument Install battery as follows 1 Remove the battery compartment cover by loosen...

Страница 7: ...hown in Figure 1 The left right and OK keys are used to launch an action indicated on the screen directly above the keys Please note that the screen is not a tough screen The labels above the keys ind...

Страница 8: ...t 3 seconds The SP 395 will turn itself off after 60 seconds without user interaction through the keys 3 HST Measurement 3 1 Measurement When powered on the SP 395 will be in the measurement read mode...

Страница 9: ...le taking readings It is recommended to use the 3mL Transfer Pipet to out the rest of the sample from the sampling cup 3 2 High Color and Turbidity Warning The SP 395 has extra channels to measure sam...

Страница 10: ...hield and then press ZERO to finish the zero calibration step If the zero calibration is successful a check mark will be displayed next to Press Zero Button If any abnormal signal is detected an error...

Страница 11: ...11 Figure 5 Figure 6...

Страница 12: ...measurement data are displayed Figure 8 The information has no use for normal operation Please provide an image of both the device information screen and the diagnosis screen when you contact Pyxis s...

Страница 13: ...ce again and then check for the flashing blue light inside the sampling cup of the SP 395 meter If the surface is not entirely clean continue to soak the SP 395 meter sampling cup for an additional 30...

Отзывы: