How to Clean SP-395
Soak the sampling cup of the SP-395 meter with handheld cleaning solution for 30 minutes. Rinse the
SP-395 sampling cup with distilled water, wipe down sampling cup with Q-tip, rinse with distilled water
once again, and then check for the flashing blue light inside the sampling cup of the SP-395 meter. If the
surface is not entirely clean, continue to soak the SP-395 meter sampling cup for an additional 30
minutes. Pyxis Lab Handheld Cleaning Solution can be purchased at our online Estore/Catalog:
https://pyxis-lab.com/product/handheld-device-cleaning-kit/
Video on how to clean handheld meters:
https://www.youtube.com/watch?v=OJDnCOjw7-M