Pyxis SP-350P Скачать руководство пользователя страница 7

3

Unpacking Instrument

Remove the instrument and accessories from the shipping container and inspect each item for any damage
that may have occurred during shipment. Verify that all items listed on the packing slip are included. If any
items are missing or damaged, please contact Pyxis Customer Service at

[email protected]

.

3.1

Standard Accessories

• Four (4) AA alkaline batteries

• Bluetooth/USB Adapter for Desktop

P/N: MA-NEB

• User Manual available online at

https://pyxis-lab.com/support/

3.2

Optional Accessories

The following optional accessories can be ordered from Pyxis Customer Service (

[email protected]

or

Pyxis eStore at

https://pyxis-lab.com/shop/

.

Table 2.

Optional Accessories

Accessory Name

Part Number

(P/N)

Pyxis Carrying Case for Handheld Devices

50725

Pyxis 10 ppm Fluorescent P 100 ppb PTSA Combined Standard — 500 mL

21055

Pyxis 10 ppm Fluorescent Polymer Calibration Standard — 500 mL

21054

Pyxis 20 ppm Fluorescent Polymer Calibration Standard — 500 mL

21053

Pyxis Handheld Cleaning Kit

SER-02

4

Installation

4.1

Battery Installation

The SP-350P is powered by four alkaline batteries. Typical battery life lasts for 3,200 measurements or 10
months. When the battery capacity is critically low, the SP-350P will display a ”LOW BATTERY” warning for
5 seconds and then automatically turn off.

*

NOTE

*

Do not use rechargeable nickel cadmium (NiCad) or lithium batteries.

Replace the batteries to resume operation of the SP-350P after the battery warning. The SP-350P will au-
tomatically turn on in the measurement mode after new batteries are installation.

The SP-350P battery compartment, shown in Figure 1, is on the back side of the instrument. Batteries are
held in place by a cover secured with two Phillips-head screws.

SP-350P User Manual

[email protected]

| +1 (866) 203-8397

4

Содержание SP-350P

Страница 1: ......

Страница 2: ...uorescent Polymer Handheld User Manual October 12 2020 Rev 2 00 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confidentia...

Страница 3: ...8 Device Information and Diagnosis 8 8 1 Sample Cell Cleanliness Check 9 8 2 Bluetooth Connection to Devices 10 8 3 Factory Reset 10 9 Use with uPyxis Mobile App 11 9 1 Download uPyxis Mobile App 11...

Страница 4: ...13 Contact Us 18 SP 350P User Manual service pyxis lab com 1 866 203 8397 1...

Страница 5: ...elivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services fo...

Страница 6: ...th the cuvette are eliminated Large color graphic screen that can be read in direct sunlight 2 Specifications Table 1 SP 350P Specifications Item Specification Part Number P N 50228 Fluorescent Polyme...

Страница 7: ...ymer 100 ppb PTSA Combined Standard 500 mL 21055 Pyxis 10 ppm Fluorescent Polymer Calibration Standard 500 mL 21054 Pyxis 20 ppm Fluorescent Polymer Calibration Standard 500 mL 21053 Pyxis Handheld Cl...

Страница 8: ...and negative terminal signs in the compartment bottom snap four new AA alkaline batteries firmly into the battery holder 4 Replace the battery compartment cover and ensure that the sealing O ring is l...

Страница 9: ...reading accuracy Before beginning a measurement use the sample water to rinse the sample cell at least three times Allow 5 10 seconds for the SP 350P to stabilize Figure 2 For a sample containing 10...

Страница 10: ...zero blank calibration 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 4 7 Rinse the sample cell three times with the desired Fluorescent Polym...

Страница 11: ...dule also shown Press Diagnosis or to launch the SYSTEM DIAGNOSIS screen where raw measurement data are displayed Figure 7 The information has no use for normal operation but instead is used for devic...

Страница 12: ...DEVICE INFORMATION screen 3 Press Diagnosis or to launch the SYSTEM DIAGNOSIS screen 4 Allow5 10secondsforthe messagein the top rightcornerof thedisplaychangefrom Starting BTLE to BTLE Started 5 Press...

Страница 13: ...r b The uPyxis Desktop App see the Use with uPyxis Desktop App section 8 3 Factory Reset Use the following steps to restore all device parameters to factory default 1 Power on the SP 350P by pressing...

Страница 14: ...h on your mobile phone Do not pair the phone Bluetooth to the SP 350P Open uPyxis Mobile App Ensure that the SP 350P is discoverable by following the steps in the Bluetooth Connection to Devices secti...

Страница 15: ...en connected the uPyxis Mobile App will default to the Setting screen From the Setting screen the user can set the Power off time and Screen off time in seconds Figure 15 SP 350P User Manual service p...

Страница 16: ...the entire duration of the update Once the update is complete the SP 350P will reboot which will disconnect the SP 350P from the uPyxis Mobile App Figure 16 10 Use with uPyxis Desktop App 10 1 Instal...

Страница 17: ...Pyxis installation 10 2 Connecting to uPyxis Desktop App Ensure that the SP 350P is discoverable by following the steps in the Bluetooth Connection to Devices section Once the app is open to find your...

Страница 18: ...e firmware file contact service pyxis lab com for these firmware files and clicking Upgrade Firmware NOTE The firmware update process takes some time and will require the SP 350P to stay within range...

Страница 19: ...onth The SP 350P will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Read Datalog deleted by clicking Delete or exported by clicking Expo...

Страница 20: ...reset the timer and the screen will re light This press will not step along the calibration process the next press needed will still be required in the sequence Always rinse the unit with clean water...

Страница 21: ...he following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to whi...

Отзывы: