Introduction

The Pyxis ST-600 Series inline sensor is uniquely designed to measure the real-time mass/mass concentra-
tion of bleach (sodium hypochlorite) by percentage as Chlorine (Cl

2

). The ST-600 Series sensor measures

the optical density of the beach solution using a near UV light source. The sensor has a built-in reference
light source and a reference light detector in addition to the main light source and main light detector. The
ST-600 Series sensor can be connected to any device that accepts an isolated or non-isolated 4–20mA input
or RS-485 Modbus. As with all Pyxis inline sensors, the ST-600 Series sensor can be wirelessly accessed via
Bluetooth/WiFi when used in conjunction with the MA-WB Bluetooth Adapter or PowerPACK Series Aux-
iliary Box and the

uPyxis®

Mobile or Desktop App. Directions on this wireless calibration capability are

included in this manual.

Primary Installation Format

The primary method is to install the sensor in-line with of the suction side of the chemical feed pump in
piping format. This method should ideally in a vertical line run to avoid chlorine gas bubble interference.
This can be done with the standard ST-001 Inline Tee Assembly (3/4” FNPT) provided with each sensor. Pyxis
also offers 2” and 3” inline tee assemblies as an accessory option if required.

Teflon Tube Installation Format

The second method is to install the sensor in-line using the unique 1/4” OD (7 mm OD) clear Teflon tubing
adapter 6 inches in length. This tubing can be passed “THROUGH” the optical channel to allow for use of
the sensor with smaller chemical feed line installations on the suction side of the chemical feed pump.

*

NOTE

*

For enhanced ambient light interference prevention, Pyxis recommends users ap-

ply black electrical tape or shrink wrap to allow sample darkening prior to and after the
optical channel as seen in Figure 3.

1.1

Main Features

The ST-600 Series sensor includes the following features:

• Can be conveniently connected to the suction or discharge side of the bleach pump in both inline Tee

(ST-001) or 1/4” Teflon tubing format with using a union; both included with sensor.

• Can be wirelessly calibrated using bleach concentration standards with the

uPyxis®

Mobile or Desktop

App via a MA-WB Bluetooth/WiFi adapter or PowerPACK Series Auxiliary Box.

• Diagnostic information (sensor tubing fouling and failure modes) can be communicated to digital

displays via Modbus RTU.

• Offers an expanded range and a temperature signal communicated via 2-channel 4–20mA outputs

and a RS-485 output, Modbus RTU.

• Easy to remove from the system for cleaning and calibration without the need for any tools.

ST-600 Series User Manual

[email protected]

| +1 (866) 203-8397

2

Содержание 50219

Страница 1: ......

Страница 2: ...Series Bleach Concentration Sensor User Manual April 28 2021 Rev 3 01 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confi...

Страница 3: ...Connecting to uPyxis Mobile App 9 5 3 Calibration Screen and Reading 10 5 4 Diagnosis Screen 11 5 5 Device Info Screen 12 6 Setup and Calibration with uPyxis Desktop App 12 6 1 Install uPyxis Desktop...

Страница 4: ...very All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services for 3...

Страница 5: ...3 inline tee assemblies as an accessory option if required Teflon Tube Installation Format The second method is to install the sensor in line using the unique 1 4 OD 7 mm OD clear Teflon tubing adapte...

Страница 6: ...readed ports Cable Length 5 ft with IP67 connectors 2 ft flying lead with IP67 adapter Power Supply 22 26 VDC 2W Dimension L Dia 6 8 1 44 inch 172 7 36 6 mm Weight 0 37 lbs 170 g Material CPVC Operati...

Страница 7: ...t P N ST 001 7 Pin Female Adapter Flying Leads Cable 2 ft P N MA 1100 Clear Teflon Tubing 1 4 OD 6 Adapter User Manual available online at https pyxis lab com support 3 2 Optional Accessories Figure 1...

Страница 8: ...bubble interfer ence Pyxis also offers 2 and 3 inline tee assemblies as an accessory option if required To properly install the ST 600 Series sensor into the ST 001 Tee Assembly follow the steps belo...

Страница 9: ...e darkening prior to and after the optical channel as seen in Figure 3 Figure 3 ST 600 Series sensor with clear Teflon tube adapter 4 3 Wiring If insufficient wattage is available from the connected c...

Страница 10: ...onnection to ST 600 Series sensor 4 5 Connecting via USB A USB RS485 adapter P N MA 485 can be used to connect a ST 600 Series sensor to a computer with the uPyxis Desktop App NOTE Using non Pyxis USB...

Страница 11: ...with uPyxis Mobile App 5 1 Download uPyxis Mobile App Download uPyxis Mobile App from Apple App Store or Google Play Figure 6 uPyxis Mobile App installation ST 600 Series User Manual service pyxis la...

Страница 12: ...1 Open uPyxis Mobile App 2 On uPyxis Mobile App pull down to refresh the list of available Pyxis devices 3 If the connection is successful the ST 600 Series and its Serial Number SN will be displayed...

Страница 13: ...t to the Calibration screen From the Calibration screen you can perform calibrations by pressing on Zero Calibration Slope Calibration and 4 20mA Span Follow the screen instructions for each calibrati...

Страница 14: ...screen you can check the diagnosis condition as well as Export Upload This feature may be used for technical support when communicating with service pyxis lab com Figure 9 ST 600 Series User Manual s...

Страница 15: ...package will download and install the Microsoft Net Framework 4 5 if not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main u...

Страница 16: ...r 2 Launch uPyxis Desktop App 3 On uPyxis Desktop App click Device Connect via USB Bluetooth or Connect via USB RS485 Fig ure 12 4 If the connection is successful the ST 600 Series sensor and its Seri...

Страница 17: ...n description for Device Name Product Name and Modbus Address then click Set to save Figure 13 6 4 Calibration Screen To calibrate the device click on Calibration On the Calibration screen there are t...

Страница 18: ...or technical support when communicating with service pyxis lab com Figure 15 7 Outputs 7 1 4 20mA Output Setup The 4 20mA output of the ST 600 sensor is scaled as Sodium Hypochlorite 4 mA 0 20 mA 16 T...

Страница 19: ...op App click 4 20mA Span found on the Calibration Screen shown in Figure 17 Figure 16 Figure 17 7 3 Communication using Modbus RTU The ST 600 Series sensor is configured as a Modbus slave device In ad...

Страница 20: ...nd cleaned if required It is sug gested that the ST 600 Series sensor be checked for fouling and cleaned calibrated on a monthly basis Heavily contaminated waters may require more frequent cleanings C...

Страница 21: ...ield the ST 600 Series sensor from direct sunlight and precipitation 9 Troubleshooting If the ST 600 Series sensor output signal is not stable and fluctuates significantly make an additional ground co...

Отзывы: