BILL OF MATERIALS FOR CAM200 (B12-2112):
Refer to Drawing #: 112-2928
Item
Part Number
Description
Quantity
1
V351
Valve Top Cap
1
2
V323
Diaphragm
1
3
V333-NPT
Fluid Section
1
4
214-14855
Plunger, stainless steel
1
5
V348
Stroke Adjust Nut
1
6
V352
Standoff
2
7
614-5898-1
Handle
1
8
V059
Spring
1
9
V300-UV
Plastic Luer Adapter
1
10
V061
Dowel Pin
1
11
V060
Set Screw
3
12
SH8-32X0.5
Socket Head Cap Screw
4
Troubleshooting
Problem
Possible Cause
Corrective Action
Valve does not
cycle
- Fluid pressure too low
- Worn diaphragm
- Increase fluid pressure above 15psi.
- Replace diaphragm
Material leaks
from valve tip
- Worn diaphragm
- Air bubble trapped in fluid body or
in dispense needle
- Replace diaphragm
- Flip valve upside down and cycle
until air bubbles are removed
Valve does not
dispense
anything
- Fluid pressure is too low
- Material cured in fluid section
- Worn diaphragm
- Increase fluid pressure above 15psi.
- Disassemble and clean valve
- Replace diaphragm
Air bubbles in
fluid
- Valve not properly purged
- Problem with fluid delivery system
- Flip valve upside down and cycle
until air bubbles are removed
- Diagnose and repair.
Dispense rate too
fast
- Fluid pressure set too high
- Stroke adjustment set too high
- Dispense tip gauge too large
- Decrease fluid pressure
- Turn stroke adjust counter-clockwise
- Replace dispense tip with smaller size
Dispense rate too
slow
- Fluid pressure set too low
- Stroke adjustment set too low
- Dispense tip gauge too small
- Increase fluid pressure
- Turn stroke adjust clockwise
- Replace dispense tip with larger size