Wuhan Raycus Fiber Laser Technologies Co., Ltd
U
ser Guide of 1000W CW laser
T1/T2 Alarm
Instruction:
Laser low / high temperature alarm occurs when the sensor in
the laser detects that the internal temperature of the laser is
abnormal.A high-temperature / low-temperature error occurs
when the temperature at the monitoring point exceeds the set
upper / lower limit.
Solutions:
High temperature alarm.
Check whether the water-cooling system is normally working,
the water temperature is set correctly, and also the water
connection is correct. When the water cooling system works
normally and the water temperature drops below 30°C, restart
the laser. If the alarm continues, please contact Raycus.
Low temperature alarm.
Check whether the actual water temperature of the water is too
low. In addition, a low ambient temperature may also cause a
low temperature alarm when the laser is cold. Please wait until
the water temperature of the water rises above 10°C and then
restrate the divice. If the alarm continues, please contact
Raycus.
Hum Alarm
Instruction:
The laser detects that the current temperature of the
water-cooled plate is lower than the current dew point
temperature of the laser. There is a risk of condensation.
Solutions:
Stop using the laser immediately. For RFL-C100 ~ RFL-C1000
lasers, please improve the working environment so that the
ambient temperature is lower than the internal temperature, and
then restart the laser (recommended to configure a separate
air-conditioned room for the laser). For the RFL-C1500S /
C2000S laser, please wait 30 minutes for the air conditioner to
run, and then restart the laser. If the alarm continues, please
contact Raycus.