PureAire Monitoring Systems, Inc.
42
Silicon Sheet
O-ring
8.
Apply a generous drop of electrolyte on the sensing electrode.
Enlarged View
Drop of
Electrolyte
Sensing
Electrode
9.
Place a new membrane on the sensor, with the bead of electrolyte between the electrode and the
membrane.
Enlarged
View
Mmebrane
Electrolyte
Bead
Sensing
Electrode
Mmebrane
Membrane
Caution: Different sensors use
membranes with either a
Green
or
Blue
tinted paper separating
each membrane.
DO NOT USE THIS PAPER
NOTE: If reusing the
old O-ring and Silicon
sheet, please dry them
thoroughly before
reinstalling.
Membrane
Color is White
NOTE:
Never touch the center
of the membrane with bare
fingers.
Membrane
8. Place the O-ring and Silicon
Sheet, (if provided) onto the sensor.