![PureAire Air Check Lite Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/pureaire/air-check-lite/air-check-lite_instruction-manual_1627208015.webp)
PureAire Monitoring Systems, Inc.
5/25/07
12
4.3.1 Routine Visual Checks
Items to check
Check for power and proper operation
Condition / status when
operating properly
Unit should be outputting a 4 mA signal when no
gas is present. On instruments with a digital
readout, display should also show 0 when no gas
is present.
Corrective actions
If analog signal (or digital display) is higher than 4
mA in a zero gas condition, adjust as required to
the environment
4.3.2 Recommended Routine Maintenance Schedule
Routine Visual Checks
Monthly
Electrolyte Replacement
Every 6 months*
Sensor O-ring and Membrane Replacement
Every 6 months*
Sensor Calibration
Every 6 months
*
Hydrogen cell can operate 1-2 years between recharges.
4.4 Loss of Power Indicator
In the event the
Air Check
Lite
loses VDC power, the 4-20 mA analog output signal drops to
0.
Содержание Air Check Lite
Страница 26: ...PureAire Monitoring Systems Inc 5 25 07 23 6 Drawings ...
Страница 28: ...PureAire Monitoring Systems Inc 5 25 07 25 DQS cell Mounted to AirCheck Lite ...
Страница 29: ...PureAire Monitoring Systems Inc 5 25 07 26 Membrane ...
Страница 30: ...PureAire Monitoring Systems Inc 5 25 07 27 24VDC 3 wire is remote up to 1000 meters ...