PureAire Monitoring Systems, Inc.
39
5.
Wipe the sensing electrode and the surrounding area with a dry paper tissue.
PAPER
TISSUE
SENSING
ELECTRODE
6.
Place a new silicone sheet (if used) and O-ring on the sensor. Make sure that the area is dry of
electrolyte.
O-RING
SILICON SHEET
Apply a drop of electrolyte on the sensing electrode.
SENSING
ELECTRODE
DROP OF
ELECTROLYTE
ENLARGED
VIEW
7.
Place a new membrane on the sensor, with the bead of electrolyte between the electrode and the
membrane.
ENLARGED
VIEW
MEMBRANE
NOTE: If reusing the old
O-ring and Silicon sheet,
please dry them thoroughly
before reinstalling.
AN sensor cell uses a
White O-ring,
p/n 1516-12W and
Silicon sheet,
p/n SC2000
Electrolyte
P/N El-4460-1
White Membrane
P/N M-4400M
Caution: A GREEN tinted
paper separates each membrane.
DO NOT USE THIS PAPER
M-4400M Membrane
Color is White
NOTE
: Never touch the
center of the membrane
with bare fingers. Oil on
your fingers can block the
membrane pores.
39