Model 90 Quantifier Sound Level Meter
Page 57
Amplitude Weighting
Q=3 (True Energy Integration)
Measurement Range
Broadband
21dB(A) to 140dB(A) Class 1
25dB(A) to 140dB(A) Class 2
143dB(C) Peak (70 to 140dB Range)
1:1 Octave Band Filters
19dB(Z) to 140dB(Z)
1:3 Octave Band Filters
14dB(Z) to 140dB(Z)
Range Steps
10-80, 20-90, 30-100, 40-110, 50-120, 60-130, 70-140
Noise Floor (Typical)
Broadband
18dB(A) Class 1, 20dB(A) Class 2
1:1 Octave Band Filters
12dB(Z) @ 1kHz 1:1 Octave Band
1:3 Octave Band Filters
7dB(Z) @ 1kHz 1:3 Octave Band
Available Measurements
The following metrics can be displayed for a recorded session and stored:
Broadband Mode
L
CPeak
L
AF
,L
AS
,L
AI
,L
CF
,L
CS
,L
CI
,L
ZF
,L
ZS
or L
ZI
(not stored)
L
AFmax
,L
ASmax
,L
AImax
,L
CFmax
,L
CSmax
,L
CImax
,L
ZFmax
,L
ZSmax
or L
ZImax
L
AFmin
,L
ASmin
,L
AImin
,L
CFmin
,L
CSmin
,L
CImin
,L
ZFmin
,L
ZSmin
or L
ZImin
L
Aeq
,L
Ceq
, or LZ
eq
L
AE
,L
CE
,or L
ZE
,L
AIeq
,L
CIeq
, or L
ZIeq
, L
AFTeq
L
0.1
to L
99.9
(five simultaneous user-selected values available)
Run time
Date and time
1 second Short L
eq
Noise Profile
The Ln data is calculated from the Sound Level data and is Time Weighted according to
the selected Time Weighting.
Filter mode
1:1 or 1:3 filter selected
Selected frequency
Filtered L
ZS
,L
ZF
or L
Z
I (not stored)
Filtered L
Zeq
(stored)
L
Aeq
, L
Ceq
, L
Zeq
(stored)
Run time
Date and time