Model 90 Quantifier Sound Level Meter
Page 40
Time Weighting
The Time Weighting of the instrument can be set to either Fast, Slow or Impulse. To set
the Time Weighting, select the Time Weighting option from the menu:
Key Press
Display
Comments
Select the required Time
Weighting and press OK to select.
The select Time Weighting is
indicated by a tick.
Press Exit to return to the main
display.
The Time Weighting applies to the following parameters:
Broadband Mode
L
AF
,L
AS
,L
AI
,L
CF
,L
CS
,L
CI
,L
ZF
,L
ZS
or L
ZI
(not stored)
L
AFmax
,L
ASmax
,L
AImax
,L
CFmax
,L
CSmax
,L
CImax
,L
ZFmax
,L
ZSmax
or L
ZImax
L
AFmin
,L
ASmin
,L
AImin
,L
CFmin
,L
CSmin
,L
CImin
,L
ZFmin
,L
ZSmin
or L
ZImin
L
0.1
to L
99.9
(five simultaneous user-selected values available)
The L
n
’s or Statistical parameters are calculated from Sound Level. Therefore, if the
Time Weighting is set to Fast, the L
n
’s will be calculated from L
AF
, and also for the Slow
and Impulse Time Weightings.
1:1 Octave Band Mode
Filtered L
ZS
,L
ZF
or L
Z
I (not stored)
1:3 Octave Band Mode
Filtered L
ZS
,L
ZF
or L
Z
I (not stored)
Frequency Weighting
The Frequency Weighting of the instrument can be set to either A, C or Z. To set the
Frequency Weighting, select the Frequency Weighting option from the menu: