For example:
1. Load Cell, Precision: 1.990 mV/V, Capacity: 10 kg
2. Load Cell, Precision: 1.980 mV/V, Capacity: 10 kg
Precision = (1.990 + 1.980)/2 = 1.985 mV/V
Capacity = 10 + 10 = 20 kg
WARNING:
IF THE
OR
VALUES ARE CHANGED,
AND
ACTIVITIES SHOULD BE REPEATED.
•
= Zero calibration page. When in the Zero calibration page, pay attention on
that scale is empty and vibration is zero. While in this page, press the button
and observe that screen flashes. Once flashing ends, process is completed. Measured
weight value is accepted as zero and saved.
•
= The page where gain calibration is made. Before making the gain
calibration, a weight which must be at the value to be measured must be put on the
scale (for example; 10 kg) and pay attention on that no vibration exists. After
pressing on the key . The weight value put on the scale is entered via the keys
and . When the key is pressed on, the value will be accepted and gain
calibration will be completed. Then device weight screen will appear. Span value
cannot be lower than %1 of the
value. Otherwise, device will alarm the error
. When the scale is not installed any weight, device will alarm the error
. If
value is selected higher than
value, device will alarm the error
. If environment has excessive amount vibration for a period more than
25 sec, device wil alarm the error
.
* Please check the Error Code's page
.