7. Maintain, Transportation and Storage
7.1 Cleaning and Disinfecting
When using alcohol wipes to disinfect the device, please air dry or clean it with clean soft cloth.
7.2 Maintain
A.
Please clean and disinfect the device before using according to the User Manual (7.1).
B.
Please recharge the battery when the screen shows
.
C.
Recharge the battery soon after the over-discharge. The device should be recharged every six
months when it is no regular used. It can extend the battery life following this guidance.
7.3 Transportation and Storage
A.
The packed device can be transported by ordinary conveyance or according to transport contract.
The device can not be transported mixed with toxic, harmful, corrosive material.
B.
The best storage environment of the device is -40 ºC ~ 60 ºC ambient temperature and not higher
than 95% relative humidity.
8. Troubleshooting
Trouble
Possible Reason
Solution
The SpO2 and Pulse Rate can not
be displayed normally
1. The finger is not properly
positioned
2.
The patient’s SpO2 is
too low
to be detected
1. Place the finger properly and
try again.
2. Try again; Go to a hospital for
a diagnosis if you are sure the
device works properly.
The SpO2 and Pulse Rate are not
displayed stably
1. The finger is not placed inside
deep enough.
2. The finger is shaking or the
patient is moving
1. Place the finger properly and
try again.
2. Calm the patient down
The device can not be turned on
1. The batteries are drained or
almost drained
2. Malfunction of the device
1. Please recharge the battery
2. Please contact the local
service center
The display suddeny turns off
1. The product will enter standby
mode without signal for 5
seconds
2. The battery is drained away or
almost drained
1. Turn on the device and insert
finger
2.Please recharge the battery
The battery can not be full
charged even after 10 hours
charging time.
The battery is broken
Please contact the local service
center
9. Key of Symbols
Symbol
Description
Symbol
Description
Refer to instruction
manual/booklet
1. The finger clip falls off ( no
finger inserted)]
2. Probe error
3. Signal inadequacy indicator
%SpO2
The pulse oxygen
saturation(%)
IP22
International Protection
bpm
Pulse rate (bpm)
WEEE (2002/96/EC)
Full-voltage
This item is compliant with
Medical Device Directive
93/42/EEC of June 14, 1993, a
directive of the European
Economic Community.
Two grid of power
One grid of power
Low-voltage
Manufacturer
Close the alarm sound
indication
Date of manufacture