Power
frequency
(50Hz)
magnetic
field
IEC 61000-4-8
3A/m
3A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment
Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The
CMS50Q1 Pulse Oximeter
is intended for use in the electromagnetic
environment specified below. The customer or the user of
CMS50Q1 Pulse
Oximeter
should assure that it is used in such an environment.
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic
environment -guidance
Radiated
RF
ICE
61000-4-
3
3V/m
80MHz
to
2.5GHz
3V/m
Portable and mobile RF
communication equipment
should be used no closer to any
part of the
CMS50Q1 Pulse
Oximeter
, including cables,
than the recommended
separation distance calculated
from the equation applicable to
the frequency of the
transmitter.