MULTIDOS
RS232 Interface
Application Constancy Check
16
D177.131.2/0
9.3 Correction Telegrams
Telegram
Description
Response
C;
x
Set/Read Air Density Correction
0: Off 1: On
C;x
CP;
xxxx.x
Set/Read nominal air pressure (500.0 <= p <= 1300.0), default: 1013.2hPa CP;xxxx.x
CT;
xx.x
Set/Read the temperature (10.0 <= T <= 40.0), default: 20/22°C
CT;xx.x
CD
Checking the correction factor temperture/air density without kQE
CD;x.xxxx
CK;
x.xxxx
Set/Read Beam Quality Factor. (0.0000 < kQE <= 2.0000), default: 1.0000
CK;x.xxxx
9.4 Calibration Factors
Telegram
Description
Response
XR;
s;cc;f
Set/Read the radiological calibration factor for channel
cc
in set
s
XR;s;cc;±m.m
mmmE+ee
XRF;
s
Set/Read the flag for set
s
for radiologically calibrated/uncalibrated
0: not calibrated, 1: calibrated
XRF;s;x
XRN;
s;name
Set/Read the name for set
s
with radiological calibration factors
XRN;s;name
XRD;s;
dd.mm
.yyyy
Set/Read the radiological calibration date for set
s
XRD;s;dd.mm
.yyyy
XRC;
s
Renew the check sum radiological calibration factor for set
s
XRC;s
9.5 Errors
Telegram
Description
Response
ERR
Checking the error status flag of MULTIDOS
Bit0:
MENU_ERROR
Bit1:
ELCAL_ERROR
Bit2-7:
not used
Bit8:
RADCAL_ERROR Set 1
Bit9:
RADCAL_ERROR Set 2
Bit10:
RADCAL_ERROR Set 3
Bit11:
RADCAL_ERROR Set 4
Bit12:
RADCAL_ERROR Set 5
Bit13-15:
not used
ERR;xxxxx