MULTIDOS
RS232 Interface
Application Dual Channel Dosemeter
D177.131.2/0
11
•####.#
Ratio of measured value of channel 2 divided by the measured value of channel 1, given in
percent. The decimal point position is fixed. Smaller numbers are right justified, with leading
spaces. Positive numbers use a space character instead of the minus sign.
If the maximum representable ratio of ±9999.9 is exceeded, •####.# will be transmitted.
If one of the measured values exceeds it limits of representation, •----.- will be transmitted.
bbbbb
Block check sequence of the response. This is a five digit number, representing a 16 bit un-
signed integer, with leading zeroes if necessary.
The block check sequence is calculated starting from the first character of the response up to
the character preceding
bbbbb
and must be equal to the value of
bbbbb
.
7.2.3 Read maximum value
The response to the DMc telegram is
DMc•m.mmE±ee
where c is the channel number (valid values are 1 or 2). The transmitted value is the maximum value of the
doserate / current measurement for the active unit (radiological or electrical) and actual measuring range of that
channel. The value is formatted in scientific notation.
7.2.4 Read Resolution
The response to the DRc telegram is
DRc•0.mmmE±ee
where c is the channel number (valid values are 1 or 2). The transmitted value is the resolution of the actual
measuring mode for the active unit (radiological or electrical) and actual measuring range of that channel. The
number of digits following the decimal point may vary from 1 to 3.
7.2.5 Measuring unit
Telegram
Response
DU
DU
u
Reads the absolute unit of measurement of the active measuring mode. Possible values of
u
:
when MULTIDOS is set to radiological units:
Gy Gy/s Gy/min Gy/h
when MULTIDOS is set to electrical units:
C A
7.3 Correction Telegrams
Telegram
Description
Response
KS
x
Set/Read the correction setting. Parameter
x
: 0 = correction off, 1 = on
KSx
KP
p
Set/Read the value of the atmospheric pressure
p
.
p
= 0500.0 … 1300.0
KPnnnn.n
KT
t
Set/Read the value of the ambient temperature
t
.
t
= 10.0 … 40.0
KTnn.n
KTR
Read the setting of the reference temperature. Parameter
x
: 0 = 20°C, 1 =
22°C
KTRx
KD
Read the calculated air density correction factor
kD
KDn.nnn
KK
ck
Set/Read the overall correction factor
k
of channel
c.
k
= 0.500 … 2.000
KKcn.nnn
7.4 Calibration Factors
Telegram
Description
Response
CR
s
F
cf
Set/Read the calibration factor
f
of channel
c
of set
s
CRsFccf
CR
s
N
name
Set/Read the name of set
s. name
is a string of 19 characters max.
CRsNname
CR
s
T
cid
Set/Read the ID of channel
c
of set
s. id
is a string of 15 characters max.
CRsTccid
CR
s
Q
q
Set/Read the measuring quantity of set
s
.
Parameter
q
: W = Dw, A = Ka, S
= Js
CRsQq
CR
s
U
x
Update (
x
= 1)/Read (no
x
) the result of the check sum protection of set
s
CRsUx
CR
s
B
bbbbb
Set/Read the block check value
bbbbb
of set
s
CRsBbbbbb