High or low results that are incorrect may have serious medical consequences. If the
result reads > (greater than) or < (less than) or results are not as expected, always
repeat the test correctly with a new unused test strip. If a test result is displayed that
is not expected, consult the following table. Any result which is inconsistent with
your medical history should immediately be reported to your physician or healthcare
provider.
Message or Issue
Probable Cause
What to do
Results are not
as expected.
Test strips improperly stored.
Repeat test, using a different vial of test strips. Run
controls, confirm that results are in range.
Batteries are defective.
Change batteries.
The analyzer was improperly stored.
Make sure analyzer was not exposed to high or low
temperatures or humidity and repeat test.
Test strip insert opening is dirty.
Clean the test strip insert opening.
MEMo Chip and test strips are not the
same lot number.
Use MEMo Chip and test strips with the same
lot number.
A displayed result
reads < (less than)
a value.
Result is below the measuring
range of the test.
If you have symptoms call a healthcare professional
immediately. Repeat the test. Run controls and
confirm that controls are in range.
A displayed result
reads > (greater than)
a value.
Result is above the measuring
range of the test.
If you have symptoms call a healthcare
professional immediately. Run test again. Run
controls and confirm that controls are in range.
Display reads ”_ _
_ _” or N/A.
Result is not available due to a
missing value from a calculation, or
a value is outside the
measuring range.
Run test again. Run controls and confirm that
controls are in range.
Unexpected Results
11
CardioChek Analyzer
Calibration Curve:
Input from MEMo Chip per test strip lot
Battery:
2 AAA 1.5 volt alkaline
Operating Temperature Range:
50-104°F (10-40°C)
Note:
The analyzer temperature must be within the test strip temperature range
to function as a system. See test strip package insert for more details.
Humidity Range:
Between 20 and 80% RH
Dimensions:
Width: 3.0 inches (7.62 cm)
Length: 5.5 inches (13.97 cm)
Height: 1.0 inches (2.54 cm)
Weight (without batteries): 4-6 oz. (113.4 - 170.1 g)