The red alarm indicator light illuminates in the event
of a problem with the high voltage power output.
Turn the power switch off after operation.
Maintenance
Periodic Maintenance
The only regular maintenance required for the
IN5130 is the periodic cleaning of the emitter points and
ionizer grills. Emitter point cleaning affects the static de-
cay ability of the ionizer and is important for maintaining
its optimal performance.
Cleaning the front grill and external chassis
Disassemble the front grill by squeezing the grill side
clips. Temporarily remove the grill. Clean it with a polyes-
ter wiper and a cleaning solution of 50% IPA (electronic-
grade isopropyl alcohol) and 50% de-ionized water. As-
semble it when it is dry. Use the same method to clean
the exterior of the unit, including the inlet grill.
Cleaning the emitter points
Contaminants will gradually accumulate on the tip
of the emitter points with continuous use. Periodic clean-
ing of the emitter points is necessary to maintain the per-
formance of the ionizer. If the emitter points are dirty,
clean them with a polyester or cotton swab and IPA. Do
not damage or loosen the emitter points.
Caution
!
The only serviceable parts inside the ionizer are
the replaceable emitter points. Any unauthor-
ized service will void the warranty and may
result in additional repair fees.