Quick Start Manual Rev 1.0
Page 18
003R-620-100
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
CAL
ZERO
⊗
Zero Calibration
Routine
Set between zero and full
scale
(Current weight displays)
<SEL>
key to start. Current weight displays.
Remove all weight.
<SEL>, <EDT>
or
<OK>
starts routine (
Z.in P
displays).
<ITM>
key to exit,
<SEL>, <EDT>
or
<OK>
to repeat routine.
SPAN
⊗
Span Calibration
Routine
Set between zero and full
scale
(Current weight displays)
<SEL>
key to start. Current weight displays.
Add test weight.
<SEL>
or
<OK>
to show calibration weight value.
Set correct weight.
<SEL>
changes position,
<EDT>
changes digit.
<ITM>
or
<OK>
starts routine (
S.in P
displays).
<ITM>
key to exit,
<SEL>, <EDT>
or
<OK>
to re-edit
calibration weight and repeat routine.
ED.LIN
⊗
Edit Linearisation
Points
Set between zero and full
scale
L1. - - -
Select
Linearisation point 1 to 5
(
L1, L2, L3, L4, L5
).
(Approx. % of full scale)
<SEL>
to step through list of points.
<OK>
to change selected point or
<ITM>
to exit.
After
<OK>
current weight displays.
Add test weight.
<OK>
to enter corrected weight.
<SEL>
changes position,
<EDT>
changes digit.
<OK>
starts routine (
L.in P
displays).
<ITM>
key to exit,
<OK>
to repeat routine.