7. Instrument Setup
7.1. Calibration
Counter
The built-in calibration counter(s) monitor the number of times
trade critical settings are altered. Refer to Trade Critical
Settings below for more information and to the
OPTION:USE
setting page17.
The table below describes when the counter(s) will increment.
Industrial or OIML:
The
Calibration Counter
increments
when trade critical settings are changed. An example of the
counter is
C.00019
.
NTEP:
Two counters display.
The
Calibration Counter
increments when trade critical
settings in the Calibration (
CAL
)
menu are changed. An
example of the counter is
C.00010
.
The
Configuration Counter
increments when other trade
critical settings (ie.
not
in the
CAL
menu) are changed. An
example of the counter is
F.00009
.
7.1.1. Trade Critical Settings
Trade critical settings can affect calibration and/or legal for
trade performance. In this document the
⊗
symbol indicates
the setting is trade critical. Each time a trade critical setting is
altered, the calibration counter will be incremented by one.
7.2. opto-LINK
For information on setting up using the opto-LINK cable refer
to opto-LINK page 6.