S8000 -100 User Manual
52
97606 Issue 1, June 2021
APPENDIX B
Address Access
Data Type Register Map Definition
Max.
Min.
Default
User configuration – Process Alarm Settings
290
R|W|F
uint16
PROCESS_ALARM_PARAMETER
11
0
0
0 = Dewpoint
1 = Temperature
2 = Pressure
3 = %rh
4 = Water content: ppm
V
5 = Water content: ppm
W
6 = Mixing ratio
7 = Absolute humidity
8 = Wetbulb
9 = Water vapour pressure
10 = Water content: percent volume
11 = Flow
291
R|W|F
uint16
PROCESS_ALARM_TYPE
4
0
1
0 = Off
1 = Over setpoint
2 = Under setpoint
3 = Inside band
4 = Outside band
292
R|W|F
float
PROCESS_ALARM_HYSTER_MS
300
0
0.2
293
R|W|F
PROCESS_ALARM_HYSTER_LS
294
R|W|F
float
PROCESS_ALARM_SETPOINT_A_MS
1999999.9
-300
-10
295
R|W|F
PROCESS_ALARM_SETPOINT_A_LS
296
R|W|F
float
PROCESS_ALARM_SETPOINT_B_MS
1999999.9
-300
0
297
R|W|F
PROCESS_ALARM_SETPOINT_B_LS
298
R|W|F
boolean
PROCESS_ALARM_OPTICS
1
0
0
0 = Off
1 = Optics warning activates process alarm
User configuration – System Alarm Settings
301
R|W|F
boolean
NOT_MEASURE_ALARM
1
0
0
0 = Off
1 = System alarm activated when not in
measurement mode
RESERVED
304
R|W|F
uint16
SET_SERIAL_TYPE
65535
0
0
737 = Activate legacy serial comms mode
(replaces Modbus)
User configuration – Ethernet
310
R
uint16
ETH_STATUS
2
0
0
0 = Error / Not fitted
1 = OK
2 = Configuring
311
R|W
uint16
ETH_IP_1
255
0
0
312
R|W
uint16
ETH_IP_2
255
0
0
313
R|W
uint16
ETH_IP_3
255
0
0
314
R|W
uint16
ETH_IP_4
255
0
0
315
R|W
uint16
ETH_SUBNET_1
255
0
0
316
R|W
uint16
ETH_SUBNET_2
255
0
0
317
R|W
uint16
ETH_SUBNET_3
255
0
0
318
R|W
uint16
ETH_SUBNET_4
255
0
0
319
R|W
uint16
ETH_GATEWAY_1
255
0
0
320
R|W
uint16
ETH_GATEWAY_2
255
0
0
321
R|W
uint16
ETH_GATEWAY_3
255
0
0
322
R|W
uint16
ETH_GATEWAY_4
255
0
0
Содержание Michell Instruments S8000 -100
Страница 1: ...S8000 100 Chilled Mirror Hygrometer User Manual 97606 Issue 1 June 2021 nbn Austria GmbH ...
Страница 51: ...S8000 100 User Manual Michell Instruments 43 APPENDIX A Appendix A Technical Specifications ...
Страница 54: ...S8000 100 User Manual 46 97606 Issue 1 June 2021 APPENDIX B Appendix B Modbus Register Map ...
Страница 63: ...S8000 100 User Manual Michell Instruments 55 APPENDIX C Appendix C Quality Recycling Warranty Information ...
Страница 65: ...S8000 100 User Manual Michell Instruments 57 APPENDIX D Appendix D Return Document Decontamination Declaration ...
Страница 67: ...S8000 100 User Manual Michell Instruments 59 Notes ...