OPERATION
3.5.3 Operational Status Display
The Operational Status display includes the following:
Mode
Reports current operational mode.
This will either be
Measure, Standby, DCC, Hold
,
Maxcool
or
Flood
.
Next Mode
Shows the time (in Hours: Minutes: Seconds) remaining until the
transition to the next mode of operation. If DCC is configured for
manual activation only, then this countdown will display --:--:--.
Process
This notification indicates whether a parameter process alarm is either
ON
or
OFF
.
The process alarm can be set on any parameter (refer to Section 3.5.10).
Film
Thickness
This figure indicates the quantity of condensate present on the mirror
on a % scale.
0% indicates condensate has not yet formed.
100% is the target level, and ±1% indicates the instrument is stable
and controlling on the dew/frost point.
TEC Drive
This symbol changes to indicate that the mirror is either being heated
or cooled. The figure indicates the % of the total available cooling or
heating power currently being used.
Logging
Indicates data logging to SD is enabled. Refer to Section
3.5.8
.
Pressure
compensation
Indicates dew point is being calculated to atmospheric pressure.
Refer to
Section 3.5.13.
Table 4
Operational Status Display
Содержание Michell Instruments S8000 -100
Страница 1: ...S8000 100 Chilled Mirror Hygrometer User Manual 97606 Issue 1 June 2021 nbn Austria GmbH ...
Страница 51: ...S8000 100 User Manual Michell Instruments 43 APPENDIX A Appendix A Technical Specifications ...
Страница 54: ...S8000 100 User Manual 46 97606 Issue 1 June 2021 APPENDIX B Appendix B Modbus Register Map ...
Страница 63: ...S8000 100 User Manual Michell Instruments 55 APPENDIX C Appendix C Quality Recycling Warranty Information ...
Страница 65: ...S8000 100 User Manual Michell Instruments 57 APPENDIX D Appendix D Return Document Decontamination Declaration ...
Страница 67: ...S8000 100 User Manual Michell Instruments 59 Notes ...