MDM300 User Manual
46
97122 Issue 16.2 June 2022
CALIBRATION
6.3
Calibration Correction Method
NOTE: Any logging procedure currently in operation must be cancelled before
the calibration procedures can be entered.
To enter the CALIBRATION Menu page, highlight
CALIBRATION
from the SET-UP Menu
by using the
or
keys and press the
Enter
key.
Entry to the CALIBRATION Menu is protected by a passcode (
4876
).
If the MDM300 has been configured for use with an external Easidew transmitter the
following page will be shown, once the passcode has been successfully entered.
Select calibration
MDM300 Calibration Correction
Easidew Offset Adjustment
Easidew One Point Adjustment
Figure 34
Calibration Page
The MDM300 Calibration look-up table is divided into three columns.
The values in the first column
POINT
represent the thirteen nominal calibration points,
spread at 10 °C intervals over the instruments operating range of -100...+20 °C
(-148...+68 °
F
)
. These points cannot be edited.
The second column
REF
represents the value of the reference dew-point standard
applied at that particular point.
The third column
MDM300
represents the reading that the MDM300 was showing at
the corresponding
REF
level input.
The values in both the
REF
and
MDM300
columns can be edited by means of the
and
keys. The
and
keys provide the means of switching between corresponding table
columns.
POINT
CALIBRATION
-100
-90
-80
-70
-60
REF
-100.0
-90.0
-80.0
-70.0
-60.0
MDM300
-100.0
-90.0
-80.0
-70.0
-60.0
Figure 35
Calibration Menu Page
After changes have been made press the
Enter
key to reach the Exit page. No changes
are implemented until the
key is pressed at the Exit page.
To exit without implementing any changes, press the
Enter
key to reach the Exit page.
Press the
key, from the Exit page, to return to the SET-UP Menu.
Содержание MICHELL Instruments MDM300
Страница 1: ...MDM300 Advanced Dew Point Hygrometer User Manual 97122 Issue 16 2 June 2022 ...
Страница 19: ...MDM300 User Manual Michell Instruments 11 INSTALLATION 1 4 3 6 5 2 Figure 6 Accessories ...
Страница 58: ...MDM300 User Manual 50 97122 Issue 16 2 June 2022 APPENDIX A Appendix A Technical Specifications ...
Страница 61: ...MDM300 User Manual Michell Instruments 53 APPENDIX B Appendix B Datalog Status Display ...
Страница 64: ...MDM300 User Manual 56 97122 Issue 16 2 June 2022 APPENDIX C Appendix C FCC Declaration ...
Страница 66: ...MDM300 User Manual 58 97122 Issue 16 2 June 2022 APPENDIX D Appendix D Quality Recycling Warranty Information ...
Страница 70: ...http www michell com www ProcessSensing com ...