PSI PlantPen/N-Pen N 110 Скачать руководство пользователя страница 12

 

 

Page | 12 

 

 

where R565 and R760 are reflectance values at indicated wavelengths. Based on a close relationship between NDGI and N-content in 
barley leaves at mid-tillering stage (Fig. 3B), the NDGI index was proposed to be a convenient tool for determination of N-content and N-
nutrition state in barley (Klem, 2008). 

Determination of NDGI index is strongly dependent on plant and leaf developmental stage as well as on environmental variability of field 
cultivation. Especially the first rapidly expanding leaf and the older leaves at the onset of senescence exhibit high variability of NDGI 
results. The most evident relationship between NDGI and N-content was reported in the second and the third youngest leaf (Klem 2008).  

 

Fig. 3 Correlation between N content in barley dry matter and reflectance at particular wavelengths indicated as Pearsons correlation coefficient along 
the reflectance spectrum (A). Relationship between NDGI index and N content in barley dry matter (B) can be interpreted as a sigmoid curve. Vertical 
bars represent 95 % confidence intervals.

 

 

 

4.2

 

N-P

EN CONFIGURATION

 

 

The N-pen N 110 is configured to quantify  reflectance of the leaf illuminated with two LED light sources (565 nm or 760 nm,  Fig. 4), 
calculate the NDGI index from average reflectance values and to predict the relative N-content in dry matter of plants or the postharvest 
grain N- content. Due to differences in leaf structure of various crops, formulas for calculation of N content were developed specifically 
for wheat, barley and corn. Each crop was tested on several varieties with similar results.  

The N-Pen is pre-set by the manufacturer for a series of 10 measurements. After completing the series, the instrument calculates average 
NDGI value and the corresponding nitrogen content in plant dry matter (%). The average value is then calculated as follows: the maximum 
and minimum values are extracted and the average is calculated of the remaining eight values. If one (or more) of the 8 values significantly 
differ from the rest (outliers), the device does not accept it (them) and asks you to repeat one or more measurements.  

The outliers typically arise with high variability of measured values, eventual leaf damage, disease infection or inproper fixing of leaf to 
measuring chamber. To exclude the outliers from the measurement the instrument uses Grubbs's test for outliers, where a t value is 
calculated as the largest absolute deviation from the sample mean. The t value is then compared with critical values stated in the table 
of Grubbs' critical values. If the calculated t value were higher than the critical value, the value is excluded due to high probability that 
the value is an outlier.  

 

Содержание PlantPen/N-Pen N 110

Страница 1: ......

Страница 2: ...he specifications of the device However deviations cannot be ruled out Therefore a complete correspondence between the manual and the real device cannot be guaranteed The information in this manual is...

Страница 3: ...on 12 4 3 Nitrogen nutrition 13 5 Getting started 14 5 1 Calibration 14 5 2 Measurement 14 6 Control menu tree 15 7 USB Connection 20 8 Bluetooth connection 21 8 1 Bluetooth pairing 21 9 FluorPen soft...

Страница 4: ...Page 4 14 Troubleshooting and Customer Support 33...

Страница 5: ...instructions in this guide Always follow corresponding manuals while working with the N Pen device or doing the maintenance It is forbidden to interfere with the hardware or software of the N Pen dev...

Страница 6: ...nts Rugged and compact N Pen can be used for plant biology research or for education Due to an inbuilt GPS module and splash proof cover is N Pen suitable for field experiments The measurement is non...

Страница 7: ...en 1 1 Windows 7 and higher Battery Li Ion rechargeable battery Capacity 2000 mAh Max charging current 0 5 A Charging via USB port PC power bank USB charger etc 70 hours typical with full operation Lo...

Страница 8: ...C ID T9J RN42 Europe ETSI EN 301 489 1 V1 8 1 ETSI EN 301 489 17 V2 1 1 ETSI EN 300 328 V1 7 1 Canada IC RSS 210 low power comm device Certification number 6514A RN42 EMC USA FCC CFR47 Part 15 subclas...

Страница 9: ...Page 9 3 DEVICE DESCRIPTION Fig 1 Device description...

Страница 10: ...manufacturer Also check the carton for any visible external damage If any damage is found notify the carrier and the manufacturer immediately The carton and all packing materials should be retained f...

Страница 11: ...region of visible spectrum Fig 2 Reflectance spectrum of green leaves roughly complements the absorption spectrum compare A and B in Fig 2 indicating that the presence of chlorophyll is critical for...

Страница 12: ...rom average reflectance values and to predict the relative N content in dry matter of plants or the postharvest grain N content Due to differences in leaf structure of various crops formulas for calcu...

Страница 13: ...is set equal to 0 as follows 1 NDGI 0 65 indicates 100 N nutrition state N 100 2 NDGI 0 25 indicates 0 N N 0 3 If 0 65 NDGI 0 25 then nutrition state is calculated as N NDGI 0 25 250 As a general rul...

Страница 14: ...s For calibration use white reflectance standard It is part of leaf clip Turn ON device hold SET button for 1 sec Close the leaf clip white standard is on bottom part of the leaf clip Go to Calibratio...

Страница 15: ...shows the Main Menu first level Sub Menus and second level Sub Menus The blue color represents the Main Menu and its Options The yellow color represents the first level Sub Menus and their Options The...

Страница 16: ...Page 16...

Страница 17: ...Page 17...

Страница 18: ...Page 18...

Страница 19: ...Page 19...

Страница 20: ...ease follow steps below in Fig 6 Fig 6 How to connect N Pen with PC A connector on the N Pen device B Portion of the USB cable with pins C E Position the cable horizontally and line up the green circl...

Страница 21: ...tooth documentation for more information Bluetooth must be switched on and be visible on both devices To pair the N Pen with another Bluetooth device such as a computer ensure that Bluetooth is switch...

Страница 22: ...Fig 8 Add a device 4 Selecting the N Pen Select PSI N Pen icon Click Next Fig 9 5 Starting the Pairing Process This step is different for old and new version of the N Pen that are equipped with dispa...

Страница 23: ...tion code is not important for the BT connection Select Next 6 Completing the N Pen Pairing Select Close Fig 12 The Bluetooth pairing is now complete and the next step is to open the program FluorPen...

Страница 24: ...the file SN txt which is included on the enclosed USB flash disk Please Note it is not possible to download data from the N Pen device without software registration 3 Switch on the N Pen and enable B...

Страница 25: ...y Erases data from the N Pen memory Online Control Online control of N Pen device Attach GPS File Used for download data from GPS module active only in version N Pen 100 MENU Setup Device ID Detects t...

Страница 26: ...opened If the box is checked the downloaded data are added to that of the current opened experiment Data GPS Visible This option is active only in older version N Pen 100 In new version the GPS data...

Страница 27: ...Control Protocols and Values These functions are not available for N Pen 9 3 DATA TRANSFER AND EXPORT 1 Perform a measurement with your N Pen 2 Click the Download icon or select Device Download 3 The...

Страница 28: ...ware update download all your data from the N Pen memory to the computer 1 Starting Update Select Setup Update Firmware From File Fig 19 2 Warning Select OK to start update Fig 20 3 Selecting bxn file...

Страница 29: ...le 4 Finishing Upload Select OK to start uploading of the update Fig 22 The bottom bar indicates the upload progress Fig 23 Fig 23 Uploading process Press OK to finish upload Fig 24 Fig 22 Data loss w...

Страница 30: ...per panel changes as shown on Fig 25 3 If the picture on the display of the device does not change then proceed to Accessories GPS Location selection in the menu and manually map the GPS by pressing S...

Страница 31: ...connect to USB port see instructions on pg 20 2 Downloading Data from the N Pen Start FluorPen program Connect N Pen device Setup Device ID Ctrl I Download measured data from the N Pen to your PC by c...

Страница 32: ...flectance and algorithms for non destructive chlorophyll assessment in higher plant leaves Journal of Plant Physiology 2003 160 271 282 Klem K Prediction of spring barley nutrition state and grain qua...

Страница 33: ...ses associated with the repair or installation of replacement parts at the customer s site The manufacturer repairs or replaces faulty instruments as quickly as possible the maximum time is one month...

Отзывы: