DIGITAL WRIST BLOOD PRESSURE MONITOR - PH 168KD
This material had been developed in interest of customer training program
by PROTON Healthcare Ltd. and downloaded from www.protonhealthcare.com
4
These specification are subject to change without notice for
purpose of improvement.
Note
This Blood Pressure Monitor complies with the EC Directive
and bears the CE mark “CE 0197”.
The quality of the device has been verified and is in line with
the provisions of the EC council directive 93/42/EEC (EN
IEC60601-1 General requirements for safety, EN IEC60601-
1-2:2001 Electromagnetic compatibility- Requirements and
tests,) dated 14 June 1993 concerning medical devices and
the EN performance standards as following:
EN 1060-1: Non-invasive sphygmomanometers - General
requirements
EN 1060-3: Non-invasive sphygmomanometers
- Supplementary requirements for
electromechanical blood pressure measuring
systems.
EN 1060-4: Non-invasive sphygmomanometers - Test
procedures to determine the overall system
accuracy of automated non-invasive
sphygmomanometers.
Specifications:
Model number
:
PH 168KD
Method of Measurement :
Oscillometric
Range of Measurement :
Pressure 0-300mmHg,
pulse 40-199 beats/minute
Accuracy Calibration
:
Pr/-3mmHg,
Pulse +/-5%
Fuzzy Pressure
:
Automatic Re-inflation
pump system
Inflation
:
Automatic inflation
(Air Pump)
Deflation of Pressure
:
Automatic air release
control valve
Exhaust
:
Automatic exhaust valve
Display
:
Liquid Crystal Digital Display
Sets of Memory
:
3x40-sets
Unit Weight
:
Approx. 148g
including batteries
Cuff Size
:
Wrist circumference
approx. 135~195mm
Operating Temperature :
+10 ̊C to + 40 ̊C, <85% R.H.
Storage Temperature
:
-20 ̊C to + 70 ̊C, <85% R.H.
Power Supply
:
2x ‘AAA’ (1.5V) Alkaline battery
Battery life
:
Approx. 250 times with
measurement
Accessories
:
Carrying case, instruction
manual, 2 pcs ‘AAA’ battery
11. Safe-keeping
•
Always keep the unit in the carrying case after its use.
•
Do not place the item directly under the sunlight, in high
temperature, or humid and dusty places.
•
Do not store in extremely low (less than – 20 °C) or high
(more than 70 °C) temperature.
12. To Prevent Malfunction
•
Do not in any way twist the wrist cuff
•
Do not start the air pressure when the wrist cuff is not
wrapped around the wrist.
•
Do not attempt to disassemble or change any parts of
the monitor including wrist cuff.
•
Do not drop or strongly shock the unit.
•
Do not operate the unit close to devices, such as
cellular or mobile phone, which are working on radio
frequency.
Warning
•
The device measures the blood pressure of adult only.
•
The device should not be operated by children in order
to avoid hazardous situatioons.
•
The unit is not supposed to be used if your wrist has
any wound or injury.
•
In case the cuff is kept pumping up and non stop,
please open the cuff at once.
Description of display marks
Display mark
Battery symbol
Measuring error
symbol
IHB symbol
ERROR
Mark appears in the
measurement condition
and flashes when pulse is
detected
Appears when the battery
voltage is excessively low
or the positions of batteries
are incorrect.
Appears when the accurate
blood pressure could not
be obtained accurately.
Appears for 1 minute
when the user was talking,
moving or shaking or
an irregular heart beat
was detected during
measurement
Condition/Cause
Corrective action
Measurements in progress.
Do not talk or move
Replace all two batteries by
new ones.
Insert the batteries at
correct positions beware of
the +/- directions.
Press “START/STOP” button
again and re-measure.
Or check cuff if wrapped
properly at the wrist or
according to instructions.
Check palm if exerting effort.
Check if talking or moving
during measurement.
Check if posture is incorrect.
Repeat the measurement.
Note that you be relaxed
for at least 5 minutes
and sit comfortably and
quietly before you restart a
measurement.
Pulse symbol