38
Protocol Systems, Inc.
3.
Place the cuff on the limb.
If possible, place it at the same
level as the heart. If above the
heart, add 1.9 mmHg to the NIBP
measurement for every inch above
the heart. If below the heart,
subtract 1.9 mmHg for every inch.
The cuff should be snug, but not
uncomfortable. The hose must not
be kinked or pinched.
Make sure the cuff tubing is
centered over the brachial artery.
If SpO
2
will also be monitored,
place the NIBP cuff on a different limb than the SpO
2
sensor to help reduce unnecessary SpO
2
alarms.
4.
Screw the hose connector onto the NIBP connector on the monitor's left side.
5.
From the Main Menu, press
NIBP
to display the NIBP Menu:
START/STOP
Starts and stops NIBP measurements. During the measurement,
you can press
STOP
(or the
Start/Stop NIBP
key at the right side of the
screen) to stop the measurement and vent the cuff.
If the Propaq CS monitor does not recognize a valid NIBP reading, it
automatically attempts another measurement while displaying a retry
message. The monitor attempts up to two retries (depending on patient
mode and settings).
AUTO/MANUAL
Switches between Automatic and Manual Mode. In Automatic Mode, the
monitor automatically takes measurements at the selected interval.
Cuff applied
evenly and
snugly with
bottom edge
of cuff one
inch above
antecubital
fossa
Possible cuff
placements for
neonates
112
58
( 85)
NIBP
18:45
S
D
(M)
mmHg
ADULT
NIBP ADULT
MODE : TURBOCUF
TIME : 4:21
SMARTCUF : ON
MON
17:05:10
PRINT FAULT
II
BR
P
1
( 96)
mmHg
(M)
TEMP
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
NIBP
( 85)
mmHg
(M)
%
°
F
122
58
MCO
2
S
p
O
2
mmHg
38
Br/m
S
D
STOP
AUTO/
MANUAL
INTERVAL
SMARTCUF
NET
OFF
ON NETWORK
T1
T2
∆
T
mV
cm
1
112
HR
BPM
mmHg
TURBOCUF
( 85)
75
50
100
150
200
17:05:01
TIME OF
READING:
(M)
122
58
S
D
12
ALLEN KIM
Manometer Bar:
Systolic, diastolic,
and mean values
are indicated as
small triangles.
The last
measurement
Measurement mode
After 16 minutes,
the NIBP display
changes as
shown below for
44 minutes.
Time remaining in
Turbocuf Mode or
Auto Mode interval
START is displayed
when no
measurement is
occurring.
This symbol
indicates the NIBP
reading was taken in
the presence of high
artifact while
monitoring ECG with
Smartcuf enabled.
Artifact can affect
accuracy. To help
reduce artifact, see
step 6 on page 39.
Smartcuf status
Start/
Stop
NIBP
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...